Cyclic Fatigue Resistance of Heat-Treated Nickel-Titanium Instruments.
Introduction: This study compared the cyclic fatigue resistance (CFR) of new instruments manufactured by heat-treated nickel-titanium wire.
Methods and Materials: Ninety-six new instruments from HyFlex CM (HF), Edge File (EF), Pro Design S (PDS/L) and Mtwo (MT) (20/0.06 and 25/0.06) ( n =12) systems were evaluated. A stainless steel device was used and time and number of cycles to fracture (NCF) were observed. Fragments were measured and fracture surface was evaluated using scanning electron microscope (SEM). ANOVA and Tukey's tests were applied with the level of significance set at 0.05.
Results: PDS 20/0.06 and PDL 25/0.06 instruments presented the highest CFR. MT 20/0.06 and MT 25/0.06 showed the lowest CFR ( P <0.05). The length of fragments was similar for 25/0.06 instruments and HF presented the highest one for 20/0.06 instruments. SEM analyses showed morphology suggestive of ductile fracture.
Conclusion: Heat treatment increased resistance to cyclic fatigue differently for each type of instrument. PDS 20/0.06 and PDL 25/0.06 present higher cyclic fatigue resistance.
Methods and Materials: Ninety-six new instruments from HyFlex CM (HF), Edge File (EF), Pro Design S (PDS/L) and Mtwo (MT) (20/0.06 and 25/0.06) ( n =12) systems were evaluated. A stainless steel device was used and time and number of cycles to fracture (NCF) were observed. Fragments were measured and fracture surface was evaluated using scanning electron microscope (SEM). ANOVA and Tukey's tests were applied with the level of significance set at 0.05.
Results: PDS 20/0.06 and PDL 25/0.06 instruments presented the highest CFR. MT 20/0.06 and MT 25/0.06 showed the lowest CFR ( P <0.05). The length of fragments was similar for 25/0.06 instruments and HF presented the highest one for 20/0.06 instruments. SEM analyses showed morphology suggestive of ductile fracture.
Conclusion: Heat treatment increased resistance to cyclic fatigue differently for each type of instrument. PDS 20/0.06 and PDL 25/0.06 present higher cyclic fatigue resistance.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app