We have located links that may give you full text access.
Reducing Treatment Errors Through Point-of-Care Glucometer Configuration.
BACKGROUND: Blood glucose (BG) testing is the most widely performed point-of-care (POC) test in a hospital setting. Multiple adverse events reported to the Food and Drug Administration (FDA) revealed that treatment decisions may be affected by information displayed on the POC glucometer's results screen. A randomized, crossover simulation study was conducted to compare two results screen configurations for ACCU-CHEK Inform II, a POC glucometer.
METHODS: Prior to the study, a heuristic evaluation of the results screen configurations and a pilot study were conducted to select the two results screen configurations for comparison. At two multicampus medical centers, 66 nurse participants experienced two computer-based simulation scenarios that asked them to interpret glucometer readings and make treatment decisions for simulated patients with 32 mg/dL BG levels and subtle symptoms of hypoglycemia. One scenario displayed a numeric value ("32 mg/dL"), and the other displayed a range abbreviation, such as "RR LO" (out of reportable range; low). Treatment errors were recorded when the participant did not treat the hypoglycemic patient with glucose or when they administered insulin.
RESULTS: When ACCU-CHEK Inform II displayed an "RR LO" reading, 10.6% of participants made a treatment error, including 6.7% of participants with prior training on the meaning of an "RR LO" reading. None of the participants made a treatment error when ACCU-CHEK Inform II displayed a "32 mg/dL" reading.
CONCLUSION: Displaying a numeric BG reading eliminated potentially life-threating treatment errors caused by confusing range abbreviations. Manufacturers should consider these findings during future research and development of POC glucometers.
METHODS: Prior to the study, a heuristic evaluation of the results screen configurations and a pilot study were conducted to select the two results screen configurations for comparison. At two multicampus medical centers, 66 nurse participants experienced two computer-based simulation scenarios that asked them to interpret glucometer readings and make treatment decisions for simulated patients with 32 mg/dL BG levels and subtle symptoms of hypoglycemia. One scenario displayed a numeric value ("32 mg/dL"), and the other displayed a range abbreviation, such as "RR LO" (out of reportable range; low). Treatment errors were recorded when the participant did not treat the hypoglycemic patient with glucose or when they administered insulin.
RESULTS: When ACCU-CHEK Inform II displayed an "RR LO" reading, 10.6% of participants made a treatment error, including 6.7% of participants with prior training on the meaning of an "RR LO" reading. None of the participants made a treatment error when ACCU-CHEK Inform II displayed a "32 mg/dL" reading.
CONCLUSION: Displaying a numeric BG reading eliminated potentially life-threating treatment errors caused by confusing range abbreviations. Manufacturers should consider these findings during future research and development of POC glucometers.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app