Transcatheter mitral valve implantation: supra-annular and subvalvular fixation techniques.

OBJECTIVES: Transcatheter mitral valved stent implantation provides an off-pump treatment option for mitral valve regurgitation, especially for secondary mitral valve regurgitation. The aim of this study was to evaluate novel fixation strategies: direct fixation (SUPRA) and subvalvular fixation (sub-VALV) to successfully implement alternative fixation methods at the mitral annulus and to reduce radial stent and apical tether forces.

METHODS: Specific concepts were developed for the supra-annular hook-shaped fixation (SUPRA) and the subvalvular fixation (sub-VALV). These prototypes were compared with the sole apical tether fixation (AP) methods. Thirty-three pigs underwent mitral valved stent implantation accompanied by standardized transoesophageal echocardiographic and haemodynamic evaluation of heart function and the stent performance 1 h after implantation. Additionally, animals were followed up for 3 months.

RESULTS: Secure deployment and correct positioning with low transvalvular gradients were achieved in all cases with mitral valved stent implantation. Nevertheless, 2 pigs died due to rhythm disturbances during dissection and pre-transcatheter mitral valve implantation. Paravalvular leakages were trace or less in prototypes with supra-annular fixation and sole apical fixation. In contrast, moderate paravalvular leakages were observed in the sub-VALV group (P < 0.001). In addition, the effect of a specific stent design on heart function was demonstrated: an increased rate of ischaemia and arrhythmia (P = 0.037) and a small left ventricular ejection fraction reduction (P < 0.05) were observed in the group with subvalvular fixation. In all but 1 case, gross evaluation demonstrated a good ingrowth of the valved stents of between 36% and 100% tissue coverage after a follow-up of 1 month or longer.

CONCLUSIONS: The low degree of paravalvular leakages in the supra-annular and apical fixation groups and the normal longitudinal function highlight the good alignment of these new mitral valved stent designs within the native anatomy. The novel, hook-shaped, supra-annular fixation elements were well tolerated in the hearts resulting in excellent health of the animals at long-term follow-up times of up to 3 months.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app