Evaluation of the 4D RADPOS dosimetry system for dose and position quality assurance of CyberKnife.

Medical Physics 2018 July 26
PURPOSE: The Synchrony respiratory motion tracking of the CyberKnife system purports to provide real-time tumor motion compensation during robotic radiosurgery. Such a complex delivery system requires thorough quality assurance. In this work, RADPOS applicability as a dose and position quality assurance tool for CyberKnife treatments is assessed quantitatively for different phantom types and breathing motions, which increase in complexity to more closely resemble clinical situations.

METHODS: Two radiotherapy treatment experiments were performed where dose and position were measured with the RADPOS probe housed within a Solid Water phantom. For the first experiment, a Solid Water breast phantom was irradiated using isocentric beam delivery while stationary or moving sinusoidally in the anterior/posterior direction. For the second experiment, a phantom consisting of a Solid Water tumor in lung equivalent material was irradiated using isocentric and non-isocentric beam delivery while either stationary or moving. The phantom movement was either sinusoidal or based on a real patient's breathing waveform. For each experiment, RADPOS dose measurements were compared to EBT3 GafChromic film dose measurements and the CyberKnife treatment planning system's (TPS) Monte Carlo and ray-tracing dose calculation algorithms. RADPOS position measurements were compared to measurements made by the CyberKnife system and to the predicted breathing motion models used by the Synchrony respiratory motion compensation.

RESULTS: For the static and dynamic (i.e., sinusoidal motion) cases of the breast experiment, RADPOS, film and the TPS agreed at the 2.0% level within 1.1 σ of estimated combined uncertainties. RADPOS position measurements were in good agreement with LED and fiducial position measurements, where the average standard deviation (SD) of the differences between any two of the three position datasets was ≤0.5 mm for all directions. For the 10 mm peak to peak amplitude sinusoidal motion of the breast experiment, the average Synchrony correlation errors were ≤0.2 mm, indicative of an accurate predictive model. For all the cases of the lung experiment, RADPOS and film measurements agreed with each other at the 2.0% level within 1.5 σ of estimated experimental uncertainties provided that the measurements were corrected for imaging dose. The measured dose for RADPOS and film were 4.0% and 3.4% higher, respectively, than the TPS for the most complex dynamic cases (i.e., irregular motion) considered for the lung experiment. Assessment of the Synchrony correlation models by RADPOS showed that model accuracy declined as motion complexity increased; the SD of the differences between RADPOS and model position data measurements was ≤0.8 mm for sinusoidal motion but increased to ≤2.6 mm for irregular patient waveform motion. These results agreed with the Synchrony correlation errors reported by the CyberKnife system.

CONCLUSIONS: RADPOS is an accurate and precise QA tool for dose and position measurements for CyberKnife deliveries with respiratory motion compensation.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app