Add like
Add dislike
Add to saved papers

A simple B 1 correction method for dynamic contrast-enhanced MRI.

The double-angle method (DAM) is commonly used as a reference standard in radiofrequency field (B 1 ) mapping studies. This study explored two aspects of DAM: (i) use of small flip angle pairs to reduce the repetition time (TR) needed for adequate longitudinal relaxation (T 1 ); and (ii) the effect of using different flip angle ratios for B 1 mapping. Results of phantom studies show that B 1 correction using small flip angle pairs ≤ 60° with TR  =  5000 ms can allow for accurate estimation of T 1 up to about 1500 ms; and that increasing the ratio of the two flip angles used for B 1 correction resulted in more accurate estimation of T 1 . These modifications allow 3-dimensional (3D) B 1 mapping to be consistently performed with the same 3D spoiled gradient echo sequence used for T 1 mapping in dynamic contrast-enhanced MRI.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app