Validity of the Actigraph GT3x and influence of the sensor positioning for the assessment of active energy expenditure during four activities of daily living in stroke subjects.
Clinical Rehabilitation 2018 July 2
OBJECTIVE: To explore the validity of the Actigraph and the influence of the placement of the sensor when estimating the active energy expenditure in four common daily activities (transfers, manual task, walking in overground and walking up and down stairs) in stroke survivors.
DESIGN: Cross-sectional study.
SETTING: Department of Physical Medicine and Rehabilitation.
SUBJECTS: A total of 46 subjects affected with stroke sequelae were wearing the Actigraph on three different locations-ankle, hip and wrist on the non-affected side-and performed four tasks: transfers, manual task, walking on flat ground and walking up and down stairs.
MAIN MEASURES: The values of active energy expenditure estimated by the Actigraph were compared to those measured by a portable breathing gas exchange analyzer Metamax3B. The accuracy and agreement between Actigraph and Metamax values were analyzed with mean bias, root mean square error, correlation coefficient and Bland-Altman plots for each task and each sensor location.
RESULTS: The mean bias between the Metamax and Actigraph placed on wrist, hip and ankle were, respectively, MD = 1.16 kcal (%MD = 3%), MD =-20.44 kcal (%MD =-58%) and MD = 17.64 kcal (%MD = 50%). The agreement with the Metamax was poor in general regardless of the sensor location and type of task ( r = 0.12-0.58).
CONCLUSION: This study found large differences and a poor agreement between the active energy expenditure as measured by the Actigraph and the Metamax according to the location of the sensor and the type of task performed by the subject.
DESIGN: Cross-sectional study.
SETTING: Department of Physical Medicine and Rehabilitation.
SUBJECTS: A total of 46 subjects affected with stroke sequelae were wearing the Actigraph on three different locations-ankle, hip and wrist on the non-affected side-and performed four tasks: transfers, manual task, walking on flat ground and walking up and down stairs.
MAIN MEASURES: The values of active energy expenditure estimated by the Actigraph were compared to those measured by a portable breathing gas exchange analyzer Metamax3B. The accuracy and agreement between Actigraph and Metamax values were analyzed with mean bias, root mean square error, correlation coefficient and Bland-Altman plots for each task and each sensor location.
RESULTS: The mean bias between the Metamax and Actigraph placed on wrist, hip and ankle were, respectively, MD = 1.16 kcal (%MD = 3%), MD =-20.44 kcal (%MD =-58%) and MD = 17.64 kcal (%MD = 50%). The agreement with the Metamax was poor in general regardless of the sensor location and type of task ( r = 0.12-0.58).
CONCLUSION: This study found large differences and a poor agreement between the active energy expenditure as measured by the Actigraph and the Metamax according to the location of the sensor and the type of task performed by the subject.
Full text links
Related Resources
Trending Papers
Heart failure with preserved ejection fraction: diagnosis, risk assessment, and treatment.Clinical Research in Cardiology : Official Journal of the German Cardiac Society 2024 April 12
Proximal versus distal diuretics in congestive heart failure.Nephrology, Dialysis, Transplantation 2024 Februrary 30
Efficacy and safety of pharmacotherapy in chronic insomnia: A review of clinical guidelines and case reports.Mental Health Clinician 2023 October
World Health Organization and International Consensus Classification of eosinophilic disorders: 2024 update on diagnosis, risk stratification, and management.American Journal of Hematology 2024 March 30
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app