Comparison of accuracy and precision between multipoint calibration, single point calibration, and relative quantification for targeted metabolomic analysis.

Targeted metabolomics requires accurate and precise quantification of candidate biomarkers, often through tandem mass spectrometric (MS/MS) analysis. Differential isotope labeling (DIL) improves mass spectrometric (MS) analysis in metabolomics by derivatizing metabolites with two isotopic forms of the same reagent. Despite its advantages, DIL-liquid chromatographic (LC)-MS/MS can result in substantial increase in workload when fully validated quantitative methods are required. To decrease the workload, we hypothesized that single point calibration or relative quantification could be used as alternative methods. Either approach will result in significant saving in resources and time. To test our hypothesis, six urinary metabolites were selected as model compounds. Urine samples were analyzed using a fully validated multipoint dansyl chloride-DIL-LC-MS/MS method. Samples were reprocessed using single point calibration and relative quantification modes. Our results demonstrated that the performance of single point calibration or relative quantification was inferior, for some metabolites, to multipoint calibration. The lower limit of quantification failed in the quantification of ethanolamine in most of participant samples using single point calibration. In addition, its precision was not acceptable in one participant during serine and ethanolamine quantification. On the other hand, relative quantification resulted in the least accurate data. In fact, none of the data generated from relative quantification for serine was comparable to that obtained from multipoint calibration. Finally, while single point calibration showed an overall acceptable performance for the majority of the model compounds, we cannot extrapolate the findings to other metabolites within the same analytical run. Analysts are advised to assess accuracy and precision for each metabolite in which single point calibration is the intended quantification mean.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app