Accidental dual humidification in intensive care units: Repeated alerts and system changes are not enough.
Journal of Critical Care 2018 October
PURPOSE: The inadvertent, simultaneous use of heat and moisture exchangers (HMEs) and heated humidifiers (HHs) can result in waterlogging of the filter and sudden ventilation tube occlusion, with potentially fatal consequences. Following an NHS England Safety Alert, a near miss and educational reminders in our institution, we introduced new guidelines to solely use HHs in the intensive care unit and HMEs only for patient transfers. No further incidents have occurred, however this solution is potentially fallible. Two years later, we sought to assess staff knowledge and likelihood of recognising this error should it occur.
MATERIALS AND METHODS: In a simulation study, a tracheally intubated and ventilated mannequin had a breathing circuit containing both a HME and a HH. Participants were asked to assess the circuit, identify errors and undertake corrective measures.
RESULTS: Only 30% (6/20) recognised and undertook corrective measures.
CONCLUSIONS: Despite educational efforts and system changes, recognition of this error remained poor. System changes may reduce the likelihood of the error occurring, but when it does, recognition may not occur. Substantial reductions or elimination of this error may be achieved through a safety-engineered fail-safe within the equipment, which alerts staff to improve recognition and prevent the mistake.
MATERIALS AND METHODS: In a simulation study, a tracheally intubated and ventilated mannequin had a breathing circuit containing both a HME and a HH. Participants were asked to assess the circuit, identify errors and undertake corrective measures.
RESULTS: Only 30% (6/20) recognised and undertook corrective measures.
CONCLUSIONS: Despite educational efforts and system changes, recognition of this error remained poor. System changes may reduce the likelihood of the error occurring, but when it does, recognition may not occur. Substantial reductions or elimination of this error may be achieved through a safety-engineered fail-safe within the equipment, which alerts staff to improve recognition and prevent the mistake.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app