Osteosynthesis of type III acromial fractures with locking compression plate, lateral clavicular plate, and reconstruction plate: a biomechanical analysis of load to failure and strain distribution.
Journal of Shoulder and Elbow Surgery 2018 November
BACKGROUND: Acromial stress fractures following reverse shoulder arthroplasty have been increasingly studied because of potential functional impairment caused by this complication. The purpose of this study was to evaluate and compare different plating techniques for a type III acromial fracture (also referred to as "scapular spine fracture") in a biomechanical setting with special regard to primary stability and modes of failure.
METHODS: Type III acromial fractures were simulated on 19 Sawbones scapulae and plated with either a lateral clavicular plate (LatCP), a locking compression plate (LCP), or a reconstruction plate (RecoP). We performed testing on 5 scapulae for each plate according to a staircase protocol (100 cycles each step): 50 N and 100 N, then increasing 100 N each step up to 800 N or until failure. The last series of mechanical tests included 3-dimensional micro-motion analysis.
RESULTS: The average force needed to cause failure of the osteosynthesis construct was 376 N for the LatCP, 506 N for the LCP, and 360 N for the RecoP. The difference between the LCP and RecoP was significant (P = .047). The average displacements of the acromion were 12.1 mm, 13.4 mm, and 11.7 mm, respectively. The spring constant was not significantly different between the plates. The LatCP showed increased strain medially, whereas the strain on the RecoP was spread more laterally. The LCP presented a balanced strain distribution, spread evenly over the fracture line.
CONCLUSIONS: In a biomechanical setting, the LCP showed superiority over the LatCP and RecoP as stabilization hardware for type III acromial fractures.
METHODS: Type III acromial fractures were simulated on 19 Sawbones scapulae and plated with either a lateral clavicular plate (LatCP), a locking compression plate (LCP), or a reconstruction plate (RecoP). We performed testing on 5 scapulae for each plate according to a staircase protocol (100 cycles each step): 50 N and 100 N, then increasing 100 N each step up to 800 N or until failure. The last series of mechanical tests included 3-dimensional micro-motion analysis.
RESULTS: The average force needed to cause failure of the osteosynthesis construct was 376 N for the LatCP, 506 N for the LCP, and 360 N for the RecoP. The difference between the LCP and RecoP was significant (P = .047). The average displacements of the acromion were 12.1 mm, 13.4 mm, and 11.7 mm, respectively. The spring constant was not significantly different between the plates. The LatCP showed increased strain medially, whereas the strain on the RecoP was spread more laterally. The LCP presented a balanced strain distribution, spread evenly over the fracture line.
CONCLUSIONS: In a biomechanical setting, the LCP showed superiority over the LatCP and RecoP as stabilization hardware for type III acromial fractures.
Full text links
Trending Papers
A Personalized Approach to the Management of Congestion in Acute Heart Failure.Heart International 2023
Potential Mechanisms of the Protective Effects of the Cardiometabolic Drugs Type-2 Sodium-Glucose Transporter Inhibitors and Glucagon-like Peptide-1 Receptor Agonists in Heart Failure.International Journal of Molecular Sciences 2024 Februrary 21
The Effect of Albumin Administration in Critically Ill Patients: A Retrospective Single-Center Analysis.Critical Care Medicine 2024 Februrary 8
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app