Validation of the Omron HBP-1300 in pregnancy for medium-arm and large-arm circumferences according to the British Hypertension Society protocol.

OBJECTIVE: The aim of this study was to validate the Omron HBP-1300 oscillometric blood pressure (BP) device according to the British Hypertension Society validation protocol, in pregnant women of both medium-arm and large-arm circumferences.

PARTICIPANTS AND METHODS: BP was measured sequentially in 72 women of any gestation requiring the use of a large-size (N=36, arm circumference ≥33 cm) or medium-size cuff (N=36, arm circumference <33 cm) alternating between a mercury sphygmomanometer and the Omron HBP-1300 device.

RESULTS: The Omron HBP-1300 is accurate in pregnancy with a mean device-observer difference of 3±6 and 1±6 mmHg for systolic BP and 2±5 and 3±6 mmHg for diastolic BP in women requiring the use of the medium and large cuff, respectively.

CONCLUSION: The Omron HBP-1300-automated BP device can be recommended for use in pregnant women with medium-arm or large-arm circumferences.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app