Add like
Add dislike
Add to saved papers

Learning curve analysis of 3D-fluoroscopy image-guided pedicle screw insertions in lumbar single-level fusion procedures.

INTRODUCTION: The implementation of 3D-navigation in the operating theater is reported to be complex, time consuming, and radiation intense. This prospective single-center cohort study was performed to objectify these assumptions by determining navigation-related learning curves in lumbar single-level posterior fusion procedures using 3D-fluoroscopy for real-time image-guided pedicle screw (PS) insertions.

MATERIALS AND METHODS: From August 2011 through July 2016, a total of 320 navigated PSs were inserted during 80 lumbar single-level posterior fusion procedures by a single surgeon without any prior experience in image-guided surgery. PS misplacements, navigation-related pre- and intraoperative time demand, and procedural 3D-radiation dose (dose-length-product, DLP) were prospectively recorded and congregated in 16 subgroups of five consecutive procedures to evaluate improving PS insertion accuracy, decreasing navigation-related time demand, and reduction of 3D-radiation dose.

RESULTS: After PS insertion and intraoperative O-arm control scanning, 11 PS modifications were performed sporadically without showing "learning curve dependencies" (PS insertion accuracies in subgroups 96.6 ± 6.3%). Average navigation-related pre-surgical time from patient positioning on the operating table to skin incision decreased from 61 ± 6 min (subgroup 1) to 28 ± 2 min (subgroup 16, p < 0.00001). Average 3D-radiation dose per surgery declined from 919 ± 225 mGycm (subgroup 1) to 66 ± 4 mGycm (subgroup 16, p < 0.0001).

CONCLUSIONS: In newly inaugurated O-arm based image-guidance, lumbar PS insertions can be performed at constantly high accuracy, even without prior experience in navigated techniques. Navigation-related time demand decreases considerably due to accelerating workflow preceding skin incision. Procedural 3D-radiation dose is reducible to a fraction (13.2%) of a lumbar diagnostic non-contrast-enhanced computed tomography scan's radiation dose.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app