Dosimetric study to assess the feasibility of intraoperative radiotherapy with electrons (ELIOT) as partial breast irradiation for patients with cardiac implantable electronic device (CIED).
Breast Cancer Research and Treatment 2018 October
PURPOSE: To report in-vivo dosimetry in the infraclavicular region, a potential site of a cardiac implantable electronic device (CIED) and to evaluate the absorbed dose from intraoperative radiotherapy with electrons (ELIOT).
METHODS: 27 non-cardiopathic breast cancer (BC) patients without CIED received quadrantectomy and ELIOT as partial breast irradiation. Before delivering ELIOT, two catheters, each containing eight thermoluminescent dosimeters (TLDs), were positioned in the infraclavicular region. TLDs internal catheter was located deep in the tumor bed while the external catheter was placed on patient's skin.
RESULTS: Data were available for 24/27 patients. The absorbed doses were referred to the dose of 21 Gy. Values measured by the external catheter were low, although statistically significant higher doses were found close to the applicator (mean values 0.26-0.49 Gy). External TLD doses in proximity of the applicator were lower than those detected by their internal counterparts. Values measured by the internal catheter TLDs varied according to the distance from the applicator while no correlation with tumor site and beam energy was found. The distance from the applicator to deliver < 2 Gy to a CIED was 2 cm, while from 2.5 cm the dose measured in all the patients became negligible.
CONCLUSIONS: This dosimetric study provided data to support the clinical use of ELIOT in BC patients having CIEDs as long as the suggested minimum safe distance of 2.5 cm is taken from the RT field in case of ELIOT single dose of 21 Gy, in the energy range of 6-10 MeV.
METHODS: 27 non-cardiopathic breast cancer (BC) patients without CIED received quadrantectomy and ELIOT as partial breast irradiation. Before delivering ELIOT, two catheters, each containing eight thermoluminescent dosimeters (TLDs), were positioned in the infraclavicular region. TLDs internal catheter was located deep in the tumor bed while the external catheter was placed on patient's skin.
RESULTS: Data were available for 24/27 patients. The absorbed doses were referred to the dose of 21 Gy. Values measured by the external catheter were low, although statistically significant higher doses were found close to the applicator (mean values 0.26-0.49 Gy). External TLD doses in proximity of the applicator were lower than those detected by their internal counterparts. Values measured by the internal catheter TLDs varied according to the distance from the applicator while no correlation with tumor site and beam energy was found. The distance from the applicator to deliver < 2 Gy to a CIED was 2 cm, while from 2.5 cm the dose measured in all the patients became negligible.
CONCLUSIONS: This dosimetric study provided data to support the clinical use of ELIOT in BC patients having CIEDs as long as the suggested minimum safe distance of 2.5 cm is taken from the RT field in case of ELIOT single dose of 21 Gy, in the energy range of 6-10 MeV.
Full text links
Related Resources
Trending Papers
Challenges in Septic Shock: From New Hemodynamics to Blood Purification Therapies.Journal of Personalized Medicine 2024 Februrary 4
Molecular Targets of Novel Therapeutics for Diabetic Kidney Disease: A New Era of Nephroprotection.International Journal of Molecular Sciences 2024 April 4
The 'Ten Commandments' for the 2023 European Society of Cardiology guidelines for the management of endocarditis.European Heart Journal 2024 April 18
A Guide to the Use of Vasopressors and Inotropes for Patients in Shock.Journal of Intensive Care Medicine 2024 April 14
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app