Biomechanical Comparison of Screw Fixation Versus a Cortical Button and Self-tensioning Suture for the Latarjet Procedure.

Background: Metal screws are traditionally used to fix the coracoid process to the glenoid. Despite stable fixation, metal screws have been associated with hardware complications. Therefore, some studies have advocated for suture button fixation during the Latarjet procedure to reduce the complications associated with screw fixation.

Purpose: To biomechanically evaluate the ultimate failure load of a cortical button and self-tensioning suture versus metal screws for coracoid graft fixation during the Latarjet procedure.

Study Design: Controlled laboratory study.

Methods: Eight matched pairs of fresh-frozen, male cadaveric shoulders (N = 16) underwent the Latarjet procedure. The shoulders of each pair were randomly assigned to 1 of 2 groups: fixation using two 3.75-mm cannulated, fully threaded metal screws or fixation using a double suture button construct. Specimens were secured in a dynamic testing machine and cyclically preconditioned from 2 to 10 N at 0.1 Hz for 10 cycles. After preconditioning, specimens were pulled to failure at a normalized displacement rate of 400% of the measured gauge length per minute. The ultimate failure load and mechanism of failure were recorded for each specimen.

Results: The mean ultimate load to failure for screw fixation (226 ± 114 N; 95% CI, 147-305 N) was not significantly different from that for suture button fixation (266 ± 73 N; 95% CI, 216-317 N) ( P = .257). The mean strain at failure for screw fixation (63% ± 21%; 95% CI, 48%-77%) was not significantly different from that for suture button fixation (86% ± 26%; 95% CI, 69%-104%) ( P = .060). The most common mechanism of failure for the screw fixation method was at the bone block drill holes, while an intramuscular rupture at the clamp-muscle interface occurred for the suture button construct.

Conclusion: The screw and suture button fixation techniques exhibited comparable biomechanical strength for coracoid bone block fixation of the Latarjet procedure.

Clinical Relevance: Metal screws have been reported to be a large contributor to intraoperative and postoperative complications. Therefore, given the results of the current study, a suture button construct may be an alternative to metal screw fixation during the Latarjet procedure. However, further clinical studies are warranted.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app