Comparison of whole-blood tacrolimus concentrations measured by different immunoassay systems.

INTRODUCTION: Different measured values for tacrolimus were obtained with different automated immunoassays. We aimed to examine the differences in the blood tacrolimus concentrations measured by the major immunoassay systems commercially available in Japan.

METHODS: Whole-blood samples from 118 patients were assayed by 3 commercial assays: chemiluminescent enzyme immunoassay (CLIA), affinity column-mediated immunoassay (ACMIA), and enzyme-multiplied immunoassay technique (EMIT). Liquid chromatography-tandem mass spectrometry (LC-MS/MS) was used for reference.

KEY FINDINGS: The correlation coefficient of immunoassay vs LC-MS/MS was excellent for ACMIA (.83) and CLIA (.81) and good for EMIT (.71). The mean error was negative for ACMIA and positive for CLIA and EMIT. The mean absolute error and root-mean-square error were almost the same for ACMIA and CLIA and lower than those for EMIT.

CONCLUSIONS: The ACMIA and CLIA yield considerably better results than the EMIT for monitoring blood tacrolimus concentrations.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app