[Administration Accuracy of Automated Infusion Device for PET Using Improved Disposable Kit].

PURPOSE: The AI-300 automated infusion device (Sumitomo Heavy Industries, Ltd., Tokyo, Japan) is subject to administration error as a function of smaller volumes of 18 F-FDG dispensed via a three-way cock supplied with a disposable kit. The present study aimed to validate the administration accuracy of the AI-300 using an improved disposable kit for quantitative positron emission tomography (PET) assessment.

METHODS: We determined administration accuracy between the improved and previous disposable kits by measuring variations in dispensed volumes and radioactive concentrations of 18 F-FDG according to the criteria of the Japanese Society of Nuclear Medicine. A reference value was generated by measuring radioactivity using a standard dose calibrator.

RESULTS: The values obtained using the previous kit deviated from the reference values by a maximum of -10.6%, and the deviation depended on dispensed volumes of 18 F-FDG<0.25 mL. In contrast, the values were relatively stable when using the improved kit with dispensed 18 F-FDG volumes < 0.25 mL. Variations in radioactive concentrations were relatively stable using the improved kit, whereas that of the previous kit was slightly unstable at high radioactive concentrations.

CONCLUSION: The administration accuracy of the AI-300 using the previous kit varied considerably according to smaller dispensed volumes, but the improved kit might alleviate this problem. The present results indicated that the improved disposal kit should be immediately implemented to eliminate uncertainty surrounding quantitative PET findings.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app