OBJECTIVES: To validate the analysis capability of RapidHIT™ 200 system for four kinds of routine forensic samples and the recyclable capability of template, template DNA and PCR products in the process of twice duplicate detection.
METHODS: The buccal swabs underwent the test twice by RapidHIT™ 200 system, and the template DNA and PCR products that arose in the system were also tested for two times. After four kinds of routine forensic samples were detected by RapidHIT™ 200 system, the follow-up tests of the template, template DNA and PCR products that arose in the system were performed.
RESULTS: The STR loci could be detected in the buccal swabs by the system for the first time. However, part of the STR loci lost during the second test. And the peak value obtained in the second test was significantly reduced than the one in the first time. The average STR loci detection rates of the template DNA and PCR products were both less than 50% in the second test, which were significantly reduced than that in the first test. In addition, the analysis capability of the system for the tissues and buccal swabs was better than that for the blood and cigarette butts. Compared with the first test, the STR loci detection rate of the tested items, template DNA and PCR products decreased with the numbers of tests.
CONCLUSIONS: RapidHIT™ 200 system is more effective in retesting buccal swabs than other samples, whereas the items, DNA template, PCR products obtained in the first and second time cannot be directly used for the further application and study of forensic medicine.
METHODS: The buccal swabs underwent the test twice by RapidHIT™ 200 system, and the template DNA and PCR products that arose in the system were also tested for two times. After four kinds of routine forensic samples were detected by RapidHIT™ 200 system, the follow-up tests of the template, template DNA and PCR products that arose in the system were performed.
RESULTS: The STR loci could be detected in the buccal swabs by the system for the first time. However, part of the STR loci lost during the second test. And the peak value obtained in the second test was significantly reduced than the one in the first time. The average STR loci detection rates of the template DNA and PCR products were both less than 50% in the second test, which were significantly reduced than that in the first test. In addition, the analysis capability of the system for the tissues and buccal swabs was better than that for the blood and cigarette butts. Compared with the first test, the STR loci detection rate of the tested items, template DNA and PCR products decreased with the numbers of tests.
CONCLUSIONS: RapidHIT™ 200 system is more effective in retesting buccal swabs than other samples, whereas the items, DNA template, PCR products obtained in the first and second time cannot be directly used for the further application and study of forensic medicine.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app