Evaluation of a new commercial method for von Willebrand factor multimeric analysis.
International Journal of Laboratory Hematology 2018 June 20
INTRODUCTION: Evaluation of von Willebrand factor (VWF) multimeric distribution is useful for subclassification of von Willebrand disease (VWD). Multimer analysis has historically been a manual, labor-intensive laboratory-developed test. The first commercial method for multimeric analysis was recently developed that utilizes a single instrument for gel electrophoresis, staining, and densitometry. The current study was undertaken to evaluate the performance characteristics of the new commercial method.
METHODS: Studies performed with the commercial method included evaluation of accuracy (method comparison), reference intervals (establishment of normal migration patterns in normal donor specimens), precision (multimer pattern reproducibility), and analytical sensitivity.
RESULTS: In the method comparison studies, concordant interpretations were obtained in 19 of 24 comparisons, including normal and abnormal specimens. The 5 specimens with discordant interpretations all involved slight differences and none were considered clinically significant. Thirty-eight normal donor specimens demonstrated normal multimer patterns. Multimer pattern reproducibility was demonstrated in normal and abnormal controls tested on each gel. In the sensitivity studies, adequate visualization of multimers was determined to require VWF protein concentrations of approximately 5%-10% of normal.
CONCLUSION: The commercial multimer method is a streamlined test that demonstrates comparable performance characteristics to our current laboratory-developed method and that provides the advantage of both electrophoresis gels and densitometry scans to aid interpretation.
METHODS: Studies performed with the commercial method included evaluation of accuracy (method comparison), reference intervals (establishment of normal migration patterns in normal donor specimens), precision (multimer pattern reproducibility), and analytical sensitivity.
RESULTS: In the method comparison studies, concordant interpretations were obtained in 19 of 24 comparisons, including normal and abnormal specimens. The 5 specimens with discordant interpretations all involved slight differences and none were considered clinically significant. Thirty-eight normal donor specimens demonstrated normal multimer patterns. Multimer pattern reproducibility was demonstrated in normal and abnormal controls tested on each gel. In the sensitivity studies, adequate visualization of multimers was determined to require VWF protein concentrations of approximately 5%-10% of normal.
CONCLUSION: The commercial multimer method is a streamlined test that demonstrates comparable performance characteristics to our current laboratory-developed method and that provides the advantage of both electrophoresis gels and densitometry scans to aid interpretation.
Full text links
Trending Papers
A Personalized Approach to the Management of Congestion in Acute Heart Failure.Heart International 2023
Potential Mechanisms of the Protective Effects of the Cardiometabolic Drugs Type-2 Sodium-Glucose Transporter Inhibitors and Glucagon-like Peptide-1 Receptor Agonists in Heart Failure.International Journal of Molecular Sciences 2024 Februrary 21
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app