Phantom-based field maps for gradient nonlinearity correction in diffusion imaging.

Gradient coils in magnetic resonance imaging do not produce perfectly linear gradient fields. For diffusion imaging, the field nonlinearities cause the amplitude and direction of the applied diffusion gradients to vary over the field of view. This leads to site- and scan-specific systematic errors in estimated diffusion parameters such as diffusivity and anisotropy, reducing reliability especially in studies that take place over multiple sites. These errors can be substantially reduced if the actual scanner-specific gradient coil magnetic fields are known. The nonlinearity of the coil fields is measured by scanner manufacturers and used internally for geometric corrections, but obtaining and using the information for a specific scanner may be impractical for many sites that operate without special-purpose local engineering and research support. We have implemented an empirical field-mapping procedure using a large phantom combined with a solid harmonic approximation to the coil fields that is simple to perform and apply. Here we describe the accuracy and precision of the approach in reproducing manufacturer gold standard field maps and in reducing spatially varying errors in quantitative diffusion imaging for a specific scanner. Before correction, median B value error ranged from 33 - 41 relative to manufacturer specification at 100 mm from isocenter; correction reduced this to 0 - 4. On-axis spatial variation in the estimated mean diffusivity of an isotropic phantom was 2.2% - 4.1% within 60 mm of isocenter before correction, 0.5% - 1.6% after. Expected fractional anisotropy in the phantom was 0; highest estimated fractional anisotropy within 60 mm of isocenter was reduced from 0.024 to 0.012 in the phase encoding direction (48% reduction) and from 0.020 to 0.006 in the frequency encoding direction (72% reduction).

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app