Catheter-Based Measurements of Absolute Coronary Blood Flow and Microvascular Resistance: Feasibility, Safety, and Reproducibility in Humans.
Circulation. Cardiovascular Interventions 2018 March
BACKGROUND: The principle of continuous thermodilution can be used to calculate absolute coronary blood flow and microvascular resistance (R). The aim of the study is to explore the safety, feasibility, and reproducibility of coronary blood flow and R measurements as measured by continuous thermodilution in humans.
METHODS AND RESULTS: Absolute coronary flow and R can be calculated by thermodilution by infusing saline at room temperature through a dedicated monorail catheter. The temperature of saline as it enters the vessel, the temperature of blood and saline mixed in the distal part of the vessel, and the distal coronary pressure were measured by a pressure/temperature sensor-tipped guidewire. The feasibility and safety of the method were tested in 135 patients who were referred for coronary angiography. No significant adverse events were observed; in 11 (8.1%) patients, bradycardia and concomitant atrioventricular block appeared transiently and were reversed immediately on interruption of the infusion. The reproducibility of measurements was tested in a subgroup of 80 patients (129 arteries). Duplicate measurements had a strong correlation both for coronary blood flow (ρ=0.841, P <0.001; intraclass correlation coefficient=0.89, P <0.001) and R (ρ=0.780, P <0.001; intraclass correlation coefficient=0.89, P <0.001). In Bland-Altman plots, there was no significant bias or asymmetry.
CONCLUSIONS: Absolute coronary blood flow (in L/min) and R (in mm Hg/L/min or Wood units) can be safely and reproducibly measured with continuous thermodilution. This approach constitutes a new opportunity for the study of the coronary microcirculation.
METHODS AND RESULTS: Absolute coronary flow and R can be calculated by thermodilution by infusing saline at room temperature through a dedicated monorail catheter. The temperature of saline as it enters the vessel, the temperature of blood and saline mixed in the distal part of the vessel, and the distal coronary pressure were measured by a pressure/temperature sensor-tipped guidewire. The feasibility and safety of the method were tested in 135 patients who were referred for coronary angiography. No significant adverse events were observed; in 11 (8.1%) patients, bradycardia and concomitant atrioventricular block appeared transiently and were reversed immediately on interruption of the infusion. The reproducibility of measurements was tested in a subgroup of 80 patients (129 arteries). Duplicate measurements had a strong correlation both for coronary blood flow (ρ=0.841, P <0.001; intraclass correlation coefficient=0.89, P <0.001) and R (ρ=0.780, P <0.001; intraclass correlation coefficient=0.89, P <0.001). In Bland-Altman plots, there was no significant bias or asymmetry.
CONCLUSIONS: Absolute coronary blood flow (in L/min) and R (in mm Hg/L/min or Wood units) can be safely and reproducibly measured with continuous thermodilution. This approach constitutes a new opportunity for the study of the coronary microcirculation.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app