Comparison of visual and automated Deki Reader interpretation of malaria rapid diagnostic tests in rural Tanzanian military health facilities.

Malaria Journal 2018 May 30
BACKGROUND: Although microscopy is a standard diagnostic tool for malaria and the gold standard, it is infrequently used because of unavailability of laboratory facilities and the absence of skilled readers in poor resource settings. Malaria rapid diagnostic tests (RDT) are currently used instead of or as an adjunct to microscopy. However, at very low parasitaemia (usually < 100 asexual parasites/µl), the test line on malaria rapid diagnostic tests can be faint and consequently hard to visualize and this may potentially affect the interpretation of the test results. Fio Corporation (Canada), developed an automated RDT reader named Deki Reader™ for automatic analysis and interpretation of rapid diagnostic tests. This study aimed to compare visual assessment and automated Deki Reader evaluations to interpret malaria rapid diagnostic tests against microscopy. Unlike in the previous studies where expert laboratory technicians interpreted the test results visually and operated the device, in this study low cadre health care workers who have not attended any formal professional training in laboratory sciences were employed.

METHODS: Finger prick blood from 1293 outpatients with fever was tested for malaria using RDT and Giemsa-stained microscopy for thick and thin blood smears. Blood samples for RDTs were processed according to manufacturers' instructions automated in the Deki Reader. Results of malaria diagnoses were compared between visual and the automated devise reading of RDT and microscopy.

RESULTS: The sensitivity of malaria rapid diagnostic test results interpreted by the Deki Reader was 94.1% and that of visual interpretation was 93.9%. The specificity of malaria rapid diagnostic test results was 71.8% and that of human interpretation was 72.0%. The positive predictive value of malaria RDT results by the Deki Reader and visual interpretation was 75.8 and 75.4%, respectively, while the negative predictive values were 92.8 and 92.4%, respectively. The accuracy of RDT as interpreted by DR and visually was 82.6 and 82.1%, respectively.

CONCLUSION: There was no significant difference in performance of RDTs interpreted by either automated DR or visually by unskilled health workers. However, despite the similarities in performance parameters, the device has proven useful because it provides stepwise guidance on processing RDT, data transfer and reporting.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app