Gripping and Anchoring Effects on the Mechanical Strengths of Orthodontic Microimplants.
Implant Dentistry 2018 June
PURPOSE: The aim of this study was to evaluate the mechanical strengths in 5 different designs of orthodontic microimplants by analyzing their configuration of structure.
MATERIALS AND METHODS: Thirty microimplants of 5 types (diameter 1.5 mm: type A, B, and C; diameter 1.3 mm: type D and E) were assessed. All microimplants were manually driven into the artificial bones at a 7-mm depth. The anchor area (AA), gripping area (GA), insertion torque (IT), Periotest value (PTV), and pullout strength (PS) were measured. Intergroup and intragroup comparisons were used to detect their significant differences.
RESULTS: In the intergroup comparison, type D had a least IT (4.5 Ncm). In the PTV analysis, type B had the largest AA (7.76 mm) and its PTV (1.6) was significantly least than the others. In the PS test, type C had the largest GA (2.40 mm) and its PS was the largest. Intragroup comparisons (IT and PS), type A, and type E presented positively significant correlation. GA revealed positive with PS, and AA showed reverse tendency with PTV.
CONCLUSION: The more AA of microimplants, the more stable they are. The more GA of microimplants, the more PS they are. Therefore, type C was better than the others because it had the largest GA and second largest AA.
MATERIALS AND METHODS: Thirty microimplants of 5 types (diameter 1.5 mm: type A, B, and C; diameter 1.3 mm: type D and E) were assessed. All microimplants were manually driven into the artificial bones at a 7-mm depth. The anchor area (AA), gripping area (GA), insertion torque (IT), Periotest value (PTV), and pullout strength (PS) were measured. Intergroup and intragroup comparisons were used to detect their significant differences.
RESULTS: In the intergroup comparison, type D had a least IT (4.5 Ncm). In the PTV analysis, type B had the largest AA (7.76 mm) and its PTV (1.6) was significantly least than the others. In the PS test, type C had the largest GA (2.40 mm) and its PS was the largest. Intragroup comparisons (IT and PS), type A, and type E presented positively significant correlation. GA revealed positive with PS, and AA showed reverse tendency with PTV.
CONCLUSION: The more AA of microimplants, the more stable they are. The more GA of microimplants, the more PS they are. Therefore, type C was better than the others because it had the largest GA and second largest AA.
Full text links
Related Resources
Trending Papers
Challenges in Septic Shock: From New Hemodynamics to Blood Purification Therapies.Journal of Personalized Medicine 2024 Februrary 4
Molecular Targets of Novel Therapeutics for Diabetic Kidney Disease: A New Era of Nephroprotection.International Journal of Molecular Sciences 2024 April 4
Perioperative echocardiographic strain analysis: what anesthesiologists should know.Canadian Journal of Anaesthesia 2024 April 11
The 'Ten Commandments' for the 2023 European Society of Cardiology guidelines for the management of endocarditis.European Heart Journal 2024 April 18
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app