Biomechanical evaluation of cable and suture cerclages for tuberosity reattachment in a 4-part proximal humeral fracture model treated with reverse shoulder arthroplasty.

BACKGROUND: Sufficient tuberosity fixation in proximal humeral fractures treated with shoulder arthroplasty is essential to gain a good clinical outcome. This biomechanical study evaluated the strength of the reattached tuberosities in reverse total shoulder arthroplasty fixed with cables or with sutures in a cerclage-like technique. Considering the mechanical advantages of flexible titanium alloy cables compared with conventional sutures for cerclage-like fixations, we hypothesized that titanium alloy cables would achieve higher fixation strengths of the tuberosities compared with heavy nonabsorbable sutures.

METHODS: A 4-part fracture was created on 8-paired proximal human humeri. The tuberosities were reduced anatomically and fixed by 2 heavy nonabsorbable sutures (suture group) or by two 1-mm titanium alloy cables (cable group) in a cerclage-like technique around the neck of the prosthesis. The humeri were placed in a custom-made test setup enabling internal and external rotation. Cyclic loading with a stepwise increasing load magnitude was applied with a material testing machine, starting with 1 Nm and increasing the load by 0.25 Nm after each 100th cycle until failure of the fixation occurred (>15° rotation of the tuberosities). Any motion of the tuberosities was measured with a 3-dimensional ultrasound motion analysis system.

RESULTS: Overall, the cable group reached 1414 ± 372 cycles, and the suture group reached 1257 ± 230 cycles until the fixations failed (P = .313). The suture group showed a significantly higher rotation of the lesser tuberosity relative to the humerus shaft axis after 200, 400, and 600 cycles compared with the cable group (P = .018-.043).

CONCLUSIONS: Tuberosities reattached with cable cerclages showed higher fixation strength and therefore less rotation compared with suture cerclages in a 4-part proximal humeral fracture model treated with reverse total shoulder arthroplasty. Whether this higher fixation strength results in higher bony ingrowth rates of the tuberosities and thus leads to a better clinical outcome needs to be investigated in further clinical studies.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app