Development of a rapid UPLC-MS/MS determination of urine sulfocysteine for diagnosis of sulfocysteinuria and molybdenum co-factor deficiencies.

Bioanalysis 2018 May 2
AIM: Molybdenum co-factor deficiencies and isolated sulfite oxidase deficiency are rare autosomal recessively inherited diseases characterized by severe psychomotor impairment, intractable seizures, dislocated lens and dysmorphic facial features. The biochemical diagnosis of these diseases requires the determination of urine sulfocysteine.

MATERIALS & METHODS: Urine sulfocysteine was quantified by an ultra-high performance liquid chromatography-MS/MS assay. The method was validated for linearity, accuracy, precision, recovery and stability.

RESULTS & CONCLUSION: Total imprecision of accuracy was less than 6%. Intra-assay and inter-assay precisions were less than 5%. The recovery was higher than 98%. The method is inexpensive, fast, accurate and has been successfully used for identifying five molybdenum co-factor deficient and six sulfite oxidase deficient patients since deployed.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app