Precise and Automatic Patient Positioning in Computed Tomography: Avatar Modeling of the Patient Surface Using a 3-Dimensional Camera.

OBJECTIVES: The aim of this study was to evaluate the accuracy of a 3-dimensional (3D) camera algorithm for automatic and individualized patient positioning based on body surface detection and to compare the results of the 3D camera with manual positioning performed by technologists in routinely obtained chest and abdomen computed tomography (CT) examinations.

MATERIALS AND METHODS: This study included data of 120 patients undergoing clinically indicated chest (n = 68) and abdomen (n = 52) CT. Fifty-two of the patients were scanned with CT using a table height manually selected by technologists; 68 patients were automatically positioned with the 3D camera, which is based on patient-specific body surface and contour detection. The ground truth table height (TGT) was defined as the table height that aligns the axial center of the patient's body region in the CT scanner isocenter. Off-centering was defined as the difference between the ground truth table height (TGT) and the actual table position used in all CT examinations. The t test was performed to determine significant differences in the vertical offset between automatic and manual positioning. The χ test was used to check whether there was a relationship between patient size and the magnitude of off-centering.

RESULTS: We found a significant improvement in patient centering (offset 5 ± 3 mm) when using the automatic positioning algorithm with the 3D camera compared with manual positioning (offset 19 ± 10 mm) performed by technologists (P < 0.005). Automatic patient positioning based on the 3D camera reduced the average offset in vertical table position from 19 mm to 7 mm for chest and from 18 mm to 4 mm for abdomen CT. The absolute maximal offset was 39 mm and 43 mm for chest and abdomen CT, respectively, when patients were positioned manually, whereas with automatic positioning using the 3D camera the offset never exceeded 15 mm. In chest CT performed with manual patient positioning, we found a significant correlation between vertical offset greater than 20 mm and patient size (body mass index, >26 kg/m, P < 0.001). In contrast, no such relationship was found for abdomen CT (P = 0.38).

CONCLUSIONS: Automatic individualized patient positioning using a 3D camera allows for accurate patient centering as compared with manual positioning, which improves radiation dose utilization.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app