Add to saved papers

Physician-Modified Thoracic Stent-Graft With Low Distal Radial Force to Prevent Distal Stent-Graft-Induced New Entry Tears in Patients With Genetic Aortic Syndromes and Aortic Dissection.

PURPOSE: To describe a novel modification technique to lower the distal radial force of a thoracic stent-graft so as to avert stent-graft-induced new entry tears (SINE) in the fragile aorta of patients with genetic aortic disease and aortic dissection.

TECHNIQUE: A commercially available thoracic stent-graft is partially deployed on a back table. The most distal Z-stent is removed, the distal fabric is marked by vascular clips, and the modified stent-graft is reloaded and deployed in the true lumen of an aortic dissection. The technique is demonstrated in 3 patients with aortic dissection related to genetic aortic diseases.

CONCLUSION: Creating a low distal radial force stent-graft is easy and can be done in a short time. Endovascular implantation appears feasible and safe.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app