The Cobalt/Chromium Ratio Provides Similar Diagnostic Value to a Low Cobalt Threshold in Predicting Adverse Local Tissue Reactions in Patients With Metal-on-Metal Hip Arthroplasty.
Journal of Arthroplasty 2018 September
BACKGROUND: A higher cobalt/chromium (Co/Cr) ratio is thought to be associated with corrosion of the trunnion and increased adverse local tissue reaction (ALTR) risk in patients treated with metal-on-metal (MoM) hip arthroplasty. The main aim of this study was to investigate the diagnostic value of Co/Cr ratio in identifying ALTR in patients treated with MoM hip resurfacing arthroplasty and total hip arthroplasty (THA).
METHODS: A total of 310 unilateral patients were included. All patients received a metal artifact reduction sequence magnetic resonance imaging to detect ALTR. Blood Co and Cr levels were measured.
RESULTS: MoM THA patients had a higher Co/Cr ratio than MoM hip resurfacing arthroplasty patients (1.7, interquartile range 1.2-3.1 vs 0.9, interquartile range 0.7-1.3, P < .001). The continuous Co/Cr ratio, continuous Co, and Co ≥ 7 ppb were predictive of ALTR when tested in receiver-operating characteristic analyses. The most sensitive and specific cutoffs were found to be 1.4 for Co/Cr ratio (sensitivity = 80.0%; specificity = 48.7%) and 2.4 ppb for Co (sensitivity = 77.8%; specificity = 46.2%). In multivariate analysis, continuous Co, Co ≥ 7 ppb, and the receiver-operating characteristic analysis-generated cutoffs were associated with ALTR prevalence.
CONCLUSION: Blood metal ions are strong, but not perfect, predictors of ALTR in patients with MoM hip arthroplasty. Higher Co/Cr ratios are associated with increased risk for ALTR in MoM THA. However, blood Co levels alone are as good a predictor of ALTR as the Co/Cr ratio. The most ALTR-sensitive cutoffs are lower than previously reported in the literature.
METHODS: A total of 310 unilateral patients were included. All patients received a metal artifact reduction sequence magnetic resonance imaging to detect ALTR. Blood Co and Cr levels were measured.
RESULTS: MoM THA patients had a higher Co/Cr ratio than MoM hip resurfacing arthroplasty patients (1.7, interquartile range 1.2-3.1 vs 0.9, interquartile range 0.7-1.3, P < .001). The continuous Co/Cr ratio, continuous Co, and Co ≥ 7 ppb were predictive of ALTR when tested in receiver-operating characteristic analyses. The most sensitive and specific cutoffs were found to be 1.4 for Co/Cr ratio (sensitivity = 80.0%; specificity = 48.7%) and 2.4 ppb for Co (sensitivity = 77.8%; specificity = 46.2%). In multivariate analysis, continuous Co, Co ≥ 7 ppb, and the receiver-operating characteristic analysis-generated cutoffs were associated with ALTR prevalence.
CONCLUSION: Blood metal ions are strong, but not perfect, predictors of ALTR in patients with MoM hip arthroplasty. Higher Co/Cr ratios are associated with increased risk for ALTR in MoM THA. However, blood Co levels alone are as good a predictor of ALTR as the Co/Cr ratio. The most ALTR-sensitive cutoffs are lower than previously reported in the literature.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app