We have located links that may give you full text access.
Sample cell for studying liquid interfaces with an in situ electric field using X-ray reflectivity and application to clay particles at oil-oil interfaces.
Journal of Synchrotron Radiation 2018 May 2
Commissioning results of a liquid sample cell for X-ray reflectivity studies with an in situ applied electrical field are presented. The cell consists of a Plexiglas container with lateral Kapton windows for air-liquid and liquid-liquid interface studies, and was constructed with grooves to accept plate electrodes on the walls parallel to the direction of the beam. Both copper and ITO plate electrodes have been used, the latter being useful for simultaneous optical studies. Commissioning tests were made at the I07 beamline of the Diamond Light Source.