First time description of early lead failure of the Linox Smart lead compared to other contemporary high-voltage leads.

PURPOSE: Early lead failure has recently been reported in ICD patients with Linox SD leads. We aimed to compare the long-term performance of the following lead model Linox Smart SD with other contemporary high-voltage leads.

METHODS: All patients receiving high-voltage leads at our center between November 2009 and May 2017 were retrospectively analyzed. Lead failure was defined as the occurrence of one or more of the following: non-physiological high-rate episodes, low- or high-voltage impedance anomalies, undersensing, or non-capture.

RESULTS: In total, 220 patients were included (Linox Smart SD, n = 113; contemporary lead, n = 107). During a median follow-up of 3.8 years (IQR 1.6-5.9 years), a total of 16 (14 in Linox Smart SD and 2 in contemporary group) lead failures occurred, mostly due to non-physiological high-rate sensing or impedance abnormalities. Lead failure incidence rates per 100 person-years were 2.9 (95% CI 1.7-4.9) and 0.6 (95% CI 0.1-2.3) for Linox Smart SD compared to contemporary leads respectively. Kaplan Meier estimates of 5-year lead failure rates were 14.0% (95% CI 8.1-23.6%) and 1.3% (95% CI 0.2-8.9%), respectively (log-rank p = 0.028). Implantation of a Linox Smart SD lead increased the risk of lead failure with a hazard ratio (HR) of 4.53 (95% CI 1.03-19.95, p = 0.046) and 4.44 (95% CI 1.00-19.77, p = 0.05) in uni- and multivariable Cox models.

CONCLUSIONS: The new Linox Smart SD lead model was associated with high failure rates and should be monitored closely to detect early signs of lead failure.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app