BACKGROUND: Muscle ultrasound (MUS) has so far not been implemented for sarcopenia assessment in clinical geriatric practice due to allegedly low reproducibility of results in the absence of standardization of procedures. However, rigorous and standardized application of this technique yields highly reproducible results. Its application, especially if integrated with clinical evaluation and comprehensive geriatric assessment, proofs very useful for rapidly obtaining information on muscle mass and architecture.
OBJECTIVE: Here, we present a standardized protocol for performing right vastus lateralis (RVL) MUS and measuring parameters of muscle size and architecture.
METHODS: RVL muscle thickness (MT), fascicle length (FL), pennation angle (PA), echo-intensity (EI) and cross-sectional area (CSA) can be assessed with this protocol. A portable instrument equipped with a 5-cm long 3-11 mHz linear probe should be used with both B-mode real-time and extended-field-of-view (EFOV) techniques. Longitudinal B-mode and transverse EFOV images should be acquired during each exam, and analyzed with NIH-ImageJ software.
CONCLUSIONS: This operative protocol represents a good compromise between the feasibility of MUS in clinical settings and the need of obtaining precise measurements of muscle parameters. Future studies should verify the reproducibility of the proposed technique, and its correlation with appendicular lean mass and parameters of muscle function.
OBJECTIVE: Here, we present a standardized protocol for performing right vastus lateralis (RVL) MUS and measuring parameters of muscle size and architecture.
METHODS: RVL muscle thickness (MT), fascicle length (FL), pennation angle (PA), echo-intensity (EI) and cross-sectional area (CSA) can be assessed with this protocol. A portable instrument equipped with a 5-cm long 3-11 mHz linear probe should be used with both B-mode real-time and extended-field-of-view (EFOV) techniques. Longitudinal B-mode and transverse EFOV images should be acquired during each exam, and analyzed with NIH-ImageJ software.
CONCLUSIONS: This operative protocol represents a good compromise between the feasibility of MUS in clinical settings and the need of obtaining precise measurements of muscle parameters. Future studies should verify the reproducibility of the proposed technique, and its correlation with appendicular lean mass and parameters of muscle function.
Full text links
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app