Development and Validation of HPLC Analytical Method for Chlorantraniliprole in Bulk and in the Formulation.
Folia Medica 2018 March 2
BACKGROUND: Chlorantraniliprole is widely used as a pesticide. It is only soluble in dimethyl formamide. However, most of the reported methods used acetonitrile and other solvents.
AIM: To develop rapid, accurate, precise, and sensitive HPLC method for chlorantraniliprole.
MATERIALS AND METHODS: 10 μG/mL of chlorantraniliprole containing solution was injected into the HPLC system and run in different solvent systems. 10 μG/mL solutions of chlorantraniprole was injected in a column with 20 μL microsyringe. The chromatogram was run for appropriate minutes with mobile phase water. The flow rate was set to 1 mL/min and detection was carried out at wavelength 270 nm. The method is validated by measuring the limit of quantitation, limit of detection, repeatability, intraday precision, inter-day precision, and accuracy. Analysis of marketed formulation, bringle, and tomato for chlorantraniprole content was also made for the developed analytical method.
RESULTS: There was linearity of chlorantraniliprole for calibration curve in between 1 and 5 μG/mL concentration. The perfect sharp peak observed in water at a retention time of 6.28 min. The limit of detection and the limit of quantitation were 0.0050 μG/mL and 0.0152 μG/mL, respectively. The study reported 99.55% repeatability, 99.49% intraday precision, 99.65% interday precision, and 99.27% accuracy.
CONCLUSION: The rapid, accurate, precise, and sensitive HPLC method for the detection of chlorantraniliprole using dimethyl formamide was developed.
AIM: To develop rapid, accurate, precise, and sensitive HPLC method for chlorantraniliprole.
MATERIALS AND METHODS: 10 μG/mL of chlorantraniliprole containing solution was injected into the HPLC system and run in different solvent systems. 10 μG/mL solutions of chlorantraniprole was injected in a column with 20 μL microsyringe. The chromatogram was run for appropriate minutes with mobile phase water. The flow rate was set to 1 mL/min and detection was carried out at wavelength 270 nm. The method is validated by measuring the limit of quantitation, limit of detection, repeatability, intraday precision, inter-day precision, and accuracy. Analysis of marketed formulation, bringle, and tomato for chlorantraniprole content was also made for the developed analytical method.
RESULTS: There was linearity of chlorantraniliprole for calibration curve in between 1 and 5 μG/mL concentration. The perfect sharp peak observed in water at a retention time of 6.28 min. The limit of detection and the limit of quantitation were 0.0050 μG/mL and 0.0152 μG/mL, respectively. The study reported 99.55% repeatability, 99.49% intraday precision, 99.65% interday precision, and 99.27% accuracy.
CONCLUSION: The rapid, accurate, precise, and sensitive HPLC method for the detection of chlorantraniliprole using dimethyl formamide was developed.
Full text links
Trending Papers
A Personalized Approach to the Management of Congestion in Acute Heart Failure.Heart International 2023
Potential Mechanisms of the Protective Effects of the Cardiometabolic Drugs Type-2 Sodium-Glucose Transporter Inhibitors and Glucagon-like Peptide-1 Receptor Agonists in Heart Failure.International Journal of Molecular Sciences 2024 Februrary 21
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app