Magnitude of cone beam CT image artifacts related to zirconium and titanium implants: impact on image quality.
Dento Maxillo Facial Radiology 2018 July
OBJECTIVES: To evaluate the magnitude of artifacts related to titanium and zirconium implants at different distances and angulations and their impact on cone beam CT(CBCT) image quality.
METHODS: CBCT images were obtained before and after the insertion of titanium and zirconium implants in a mandible on different CBCT units: Picasso Trio, ProMax 3D and 3D Accuitomo 80. Artifact was assessed by measuring the standard deviation (SD) of gray values and contrast-to-noise ratio (CNR) of 11 regions of interest (ROIs) at different distances (1.5 cm, 2.5 cm and 3.5 cm) and angulations (65°, 90°, 115° and 140°) from implant region.
RESULTS: For titanium images, SD values did not differ from those of images without implant in all ROIs; however, some effect occurred in Picasso images as higher values were observed in ROIs closer to the implant (p < 0.05). Zirconium images showed higher SD values than the others in some ROIs for Picasso and ProMax (p < 0.05). In ProMax, the difference was observed even in the farthest ROIs from the implant. CNR values were not influenced by the ROI in Picasso, but presented lower values in ROIs closer to the zirconium implant for ProMax and Accuitomo.
CONCLUSIONS: The quantity and magnitude of artifacts in CBCT are influenced by the type of implant and CBCT unit. Although they are more pronounced in regions closer to the implant and located at 90° in relation to the mandibular long axis, they can reach as far as 3.5 cm from the artifact-generator object.
METHODS: CBCT images were obtained before and after the insertion of titanium and zirconium implants in a mandible on different CBCT units: Picasso Trio, ProMax 3D and 3D Accuitomo 80. Artifact was assessed by measuring the standard deviation (SD) of gray values and contrast-to-noise ratio (CNR) of 11 regions of interest (ROIs) at different distances (1.5 cm, 2.5 cm and 3.5 cm) and angulations (65°, 90°, 115° and 140°) from implant region.
RESULTS: For titanium images, SD values did not differ from those of images without implant in all ROIs; however, some effect occurred in Picasso images as higher values were observed in ROIs closer to the implant (p < 0.05). Zirconium images showed higher SD values than the others in some ROIs for Picasso and ProMax (p < 0.05). In ProMax, the difference was observed even in the farthest ROIs from the implant. CNR values were not influenced by the ROI in Picasso, but presented lower values in ROIs closer to the zirconium implant for ProMax and Accuitomo.
CONCLUSIONS: The quantity and magnitude of artifacts in CBCT are influenced by the type of implant and CBCT unit. Although they are more pronounced in regions closer to the implant and located at 90° in relation to the mandibular long axis, they can reach as far as 3.5 cm from the artifact-generator object.
Full text links
Trending Papers
A Personalized Approach to the Management of Congestion in Acute Heart Failure.Heart International 2023
Potential Mechanisms of the Protective Effects of the Cardiometabolic Drugs Type-2 Sodium-Glucose Transporter Inhibitors and Glucagon-like Peptide-1 Receptor Agonists in Heart Failure.International Journal of Molecular Sciences 2024 Februrary 21
The Effect of Albumin Administration in Critically Ill Patients: A Retrospective Single-Center Analysis.Critical Care Medicine 2024 Februrary 8
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app