Percutaneous upper extremity fracture fixation using a novel glass-based adhesive.
Journal of Orthopaedics 2018 March
Objective: To develop a surgical technique for percutaneous upper extremity fracture fixation using a novel glass-based adhesive.
Methods: Three intact upper extremity cadaveric specimens with undisturbed soft tissues were obtained. Two were used to model a wrist fracture, and the third to model a proximal humerus fracture. Fractures were produced using a small osteotome in a percutaneous fashion. Banna Bone Adhesive (BBA) was delivered to the fracture site percutaneously using a 16 gauge needle under bi-planar fluoroscopic guidance. After setting of the adhesive, the specimens were dissected to qualitatively assess BBA delivery and placement.
Results: The adhesive could readily be delivered through the 16 gauge needle with an appropriate amount of pressure applied to the syringe. Using the fluoroscope, the adhesive could be seen to flow into the fracture site with minimal extravagation into the surrounding soft tissues. Successful bonding of the fracture fragments was observed.
Conclusions: Percutaneous delivery of BBA into a fracture of the distal radius and proximal humerus may be a feasible fracture fixation technique. Biomechanical testing and animal model testing are required to further develop this procedure.
Methods: Three intact upper extremity cadaveric specimens with undisturbed soft tissues were obtained. Two were used to model a wrist fracture, and the third to model a proximal humerus fracture. Fractures were produced using a small osteotome in a percutaneous fashion. Banna Bone Adhesive (BBA) was delivered to the fracture site percutaneously using a 16 gauge needle under bi-planar fluoroscopic guidance. After setting of the adhesive, the specimens were dissected to qualitatively assess BBA delivery and placement.
Results: The adhesive could readily be delivered through the 16 gauge needle with an appropriate amount of pressure applied to the syringe. Using the fluoroscope, the adhesive could be seen to flow into the fracture site with minimal extravagation into the surrounding soft tissues. Successful bonding of the fracture fragments was observed.
Conclusions: Percutaneous delivery of BBA into a fracture of the distal radius and proximal humerus may be a feasible fracture fixation technique. Biomechanical testing and animal model testing are required to further develop this procedure.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app