Effect of cleaning methods on retentive values of saliva-contaminated implant-supported zirconia copings.

OBJECTIVE: The aim of this study was to evaluate the effect of different cleaning regimens on retentive strength of saliva-contaminated implant-supported zirconia copings.

MATERIAL AND METHODS: Seventy solid titanium abutments with 5.5 mm height (Straumann) were attached to the regular neck implant analogs (Straumann). The abutment-analog complex was mounted vertically in an autopolymerized T-shaped acrylic resin block. Seventy zirconia copings with an occlusal loop were made. The copings were contaminated with fresh human saliva for 1 min (except the control group). Afterward, the specimens were washed with water spray for 15 s and dried for 15 s. The copings were divided into seven groups according to cleaning methods (n = 10). Group 1: no contamination (control group); Group 2: water spray rinsing; Group 3: airborne particle abrasion; Group 4: immersion in 96% isopropanol; Group 5: applying Ivoclean (Ivoclar Vivadent); Group 6: applying 1% sodium hypochlorite; and Group 7: applying Monobond Plus (Ivoclar Vivadent). The copings were luted with a resin luting agent (RelyX Unicem; 3M ESPE). After 5,000 thermal cycles, the retentive values of the restorations were tested using universal testing machine. The dislodging forces were analyzed using one-way analysis of variance and the Tukey's HSD tests (α = 0.05).

RESULTS: The copings, which were cleaned with Ivoclean and 96% isopropanol, showed the highest retentive values (204.79 ± 44.49 and 232.65 ± 71.36, respectively). There was a significant difference between the studied groups (F = 2.735; p = .02). Tukey's HSD showed that there was no significant difference between the different cleaning procedures and control group except water rinsing group (p = .14). The lowest retentive value was related to the saliva-contaminated group, which were only cleaned with water rinsing method.

CONCLUSION: The retentive values of zirconia-based restorations were adversely affected by saliva contaminations. These restorations can be cleaned by Ivoclean, 96% isopropanol, airborne particle abrasion, 1% sodium hypochlorite, or applying Monobond Plus before luting procedures.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app