Add like
Add dislike
Add to saved papers

Effects of Impella on Coronary Perfusion in Patients With Critical Coronary Artery Stenosis.

BACKGROUND: Mechanical circulatory support devices are used to maintain hemodynamic stability during high-risk percutaneous coronary interventions. Little is known on the effects of such devices on coronary hemodynamics in patients with significant coronary stenosis. We sought to investigate whether mechanical circulatory support in the form of Impella (Abiomed Inc, Danvers, MA) can improve coronary hemodynamics in the presence of a critical coronary stenosis.

METHODS AND RESULTS: We examined coronary perfusion pressures and coronary pressure distal to a critical stenosis using a coronary pressure wire in 11 patients (12 coronary lesions) undergoing high-risk percutaneous coronary interventions with the use of mechanical circulatory support. Systemic, ventricular, and coronary hemodynamics were obtained at both minimum and maximum support levels before high-risk percutaneous coronary interventions. All patients had obstructive lesions with angiographically estimated diameter stenosis between 70% and 99% and distal coronary artery pressure to aortic pressure ratios between 0.44 and 0.88. When compared with minimum support, maximum support resulted in a decrease in the left ventricular end-diastolic pressure (27.3±8.6 versus 21.5±5.2 mm Hg; P =0.002) and increases in the mean systemic blood pressure (77.6±13.5 versus 88.2±12.2 mm Hg; P <0.001) and mean distal coronary pressure (51.8±20.2 versus 60.8±18.1 mm Hg; P <0.001). Effective coronary perfusion pressure (mean aortic pressure-left ventricular end-diastolic pressure) significantly increased with maximum support (49.8±15.7 versus 67.2±13.6 mm Hg; P <0.001). Diastolic perfusion pressure (diastolic blood pressure-left ventricular end-diastolic pressure) also significantly increased with maximum support (32.9±13.4 versus 52.0±11.6 mm Hg; P <0.001).

CONCLUSIONS: Mechanical circulatory support with Impella can improve distal coronary pressure and coronary perfusion pressures in the presence of critical coronary stenosis.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app