Accuracy analysis of computer-assisted surgery for femoral trochanteric fracture using a fluoroscopic navigation system: Stryker ADAPT ® system.
Injury 2018 June
PURPOSE: ADAPT is a fluoroscopic computer-assisted surgery system which intraoperatively shows the distance from the tip of the screw to the surface of the femoral head, tip-to-head-surface distance (TSD), and the tip-apex distance (TAD) advocated by Baumgaertner et al. The study evaluated the accuracy of ADAPT.
PATIENTS AND METHODS: A total of 55 patients operated with ADAPT between August 2016 and March 2017 were included as subjects. TSD and TAD were measured postoperatively using computed tomography (CT) and X-rays. The intraclass correlation coefficient (ICC) was checked in advance. The error was defined as the difference between postoperative and intraoperative measurement values of ADAPT. Summary statistics, root mean square errors (RMSEs), and correlations were evaluated.
RESULTS: ICC was 0.94 [95% CI: 0.90-0.96] in TSD and 0.99 [95% CI: 0.98-0.99] in TAD. The error was -0.35 mm (-1.83 mm to 1.12 mm) in TSD and +0.63 mm (-5.65 mm to 4.59 mm) in TAD. RMSE was 0.63 mm in TSD and 1.53 mm in TAD. Pearson's correlation coefficient was 0.79 [95% CI: 0.66-0.87] in TSD and 0.83 [95% CI: 0.72-0.89] in TAD. There were no adverse events with ADAPT use.
CONCLUSION: ADAPT is highly accurate and useful in guiding surgeons in properly positioning the screws.
PATIENTS AND METHODS: A total of 55 patients operated with ADAPT between August 2016 and March 2017 were included as subjects. TSD and TAD were measured postoperatively using computed tomography (CT) and X-rays. The intraclass correlation coefficient (ICC) was checked in advance. The error was defined as the difference between postoperative and intraoperative measurement values of ADAPT. Summary statistics, root mean square errors (RMSEs), and correlations were evaluated.
RESULTS: ICC was 0.94 [95% CI: 0.90-0.96] in TSD and 0.99 [95% CI: 0.98-0.99] in TAD. The error was -0.35 mm (-1.83 mm to 1.12 mm) in TSD and +0.63 mm (-5.65 mm to 4.59 mm) in TAD. RMSE was 0.63 mm in TSD and 1.53 mm in TAD. Pearson's correlation coefficient was 0.79 [95% CI: 0.66-0.87] in TSD and 0.83 [95% CI: 0.72-0.89] in TAD. There were no adverse events with ADAPT use.
CONCLUSION: ADAPT is highly accurate and useful in guiding surgeons in properly positioning the screws.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app