Histological and Thermometric Examination of Soft Tissue De-Epithelialization Using Digitally Controlled Er:YAG Laser Handpiece: An Ex Vivo Study.

OBJECTIVE: The purpose of this study was histological and thermometric examination of soft tissue de-epithelialization using digitally controlled laser handpiece (DCLH) - X-Runner.

BACKGROUND DATA: Commonly used techniques for de-epithelialization include scalpel, abrasion with diamond bur, or a combination of the two. Despite being simple, inexpensive and effective, these techniques are invasive and may produce unwanted side effects. It is important to look for alternative techniques using novel tools, which are minimally invasive and effective.

MATERIALS AND METHODS: 114 porcine samples sized 6 × 6 mm were collected from the attached gingiva (AG) of the alveolar process of the mandible using 15C scalpel blade. The samples were irradiated by means of Er:YAG laser (LightWalker, Fotona, Slovenia), using X-Runner and HO2 handpieces at different parameters; 80, 100, and 140 mJ/20 Hz in time of 6 or 16 sec, respectively. The temperature was measured with a K-type thermocouple. For the histopathological analysis of efficiency of epithelium removal and thermal injury, 3 random samples were de-epithelialized with an HO2 handpiece, and 9 random samples with an X-Runner handpiece with different parameters. For the samples irradiated with DCLH, we have used three different settings, which resulted in removing 1 to 3 layers of the soft tissue. The efficiency of epithelium removal and the rise of temperature were analyzed.

RESULTS: DCLH has induced significantly lower temperature increase compared with HO2 at each energy to frequency ratio. The histological examination revealed total epithelium removal when HO2 handpiece was used at 100 and 140 mJ/20 Hz and when DCLH was used for two- and threefold lasing at 80, 100, and 140 mJ/20 Hz.

CONCLUSIONS: Er:YAG laser with DCLH handpiece may be an efficient tool in epithelium removal without excessive thermal damage.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app