Comparison of central corneal thickness measurements by ultrasound pachymetry and 2 new devices, Tonoref III and RS-3000.
International Ophthalmology 2018 March 22
PURPOSE: There is currently little evidence assessing the repeatability and accuracy of central corneal thickness measurements using the devices Tonoref III and RS-3000. This study aims to compare these devices against measurements by ultrasound pachymetry.
METHODS: Central corneal thicknesses were measured on 50 eyes. Measurements from two non-contact devices-Tonoref III (NIDEK CO., LTD, Gamagori, Japan) and RS-3000 (NIDEK CO., LTD, Gamagori, Japan)-were compared against ultrasound pachymetry, the gold standard. Ultrasound measurements were obtained using a 'Pachmate' device (DGH Technology, Inc, Exton, PA, USA). Repeatability was defined as the value that the difference between two consecutive measurements falls below 95% of the time. The within-subject standard deviation and repeatability values were calculated for Pachmate and Tonoref III by one-way ANOVA. Repeatability of RS-3000 was determined by nonparametric analysis. Agreement between Tonoref III and Pachmate was assessed by a Bland-Altman plot. Agreement between RS-3000 and Pachmate was assessed by nonparametric analysis.
RESULTS: The Pachmate, Tonoref III and RS-3000 had repeatability values of 16, 7.4 and 5 µm, respectively. The mean difference between Tonoref III and Pachmate was - 15 µm (95% LoAs - 31 to + 0 µm). The median value for the difference between RS-3000 and Pachmate was - 4 µm (95% of values within - 24 and + 4 µm).
CONCLUSION: The Tonoref III and RS-3000 showed good repeatability when compared to ultrasound pachymetry. However, neither instrument agreed interchangeably with CCT measurements by ultrasound pachymetry. Practitioners should determine whether the level of agreement is sufficient to meet their clinical needs.
METHODS: Central corneal thicknesses were measured on 50 eyes. Measurements from two non-contact devices-Tonoref III (NIDEK CO., LTD, Gamagori, Japan) and RS-3000 (NIDEK CO., LTD, Gamagori, Japan)-were compared against ultrasound pachymetry, the gold standard. Ultrasound measurements were obtained using a 'Pachmate' device (DGH Technology, Inc, Exton, PA, USA). Repeatability was defined as the value that the difference between two consecutive measurements falls below 95% of the time. The within-subject standard deviation and repeatability values were calculated for Pachmate and Tonoref III by one-way ANOVA. Repeatability of RS-3000 was determined by nonparametric analysis. Agreement between Tonoref III and Pachmate was assessed by a Bland-Altman plot. Agreement between RS-3000 and Pachmate was assessed by nonparametric analysis.
RESULTS: The Pachmate, Tonoref III and RS-3000 had repeatability values of 16, 7.4 and 5 µm, respectively. The mean difference between Tonoref III and Pachmate was - 15 µm (95% LoAs - 31 to + 0 µm). The median value for the difference between RS-3000 and Pachmate was - 4 µm (95% of values within - 24 and + 4 µm).
CONCLUSION: The Tonoref III and RS-3000 showed good repeatability when compared to ultrasound pachymetry. However, neither instrument agreed interchangeably with CCT measurements by ultrasound pachymetry. Practitioners should determine whether the level of agreement is sufficient to meet their clinical needs.
Full text links
Related Resources
Trending Papers
Challenges in Septic Shock: From New Hemodynamics to Blood Purification Therapies.Journal of Personalized Medicine 2024 Februrary 4
Molecular Targets of Novel Therapeutics for Diabetic Kidney Disease: A New Era of Nephroprotection.International Journal of Molecular Sciences 2024 April 4
The 'Ten Commandments' for the 2023 European Society of Cardiology guidelines for the management of endocarditis.European Heart Journal 2024 April 18
A Guide to the Use of Vasopressors and Inotropes for Patients in Shock.Journal of Intensive Care Medicine 2024 April 14
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app