PURPOSE: To report the intricacies of managing complications that arose out of cosmetic iris implants (BrightOcular) placement.
DESIGN: Interventional case report.
METHODS: A thirty-year-old gentleman presented with complaint of progressive loss of vision after having cosmetic iris implant surgery to change his eye colour. He then developed raised intraocular pressures and had a right eye trabeculectomy conducted with the implants in situ. Subsequently, he had implant removal surgery because of persistent implant-associated complications. The vision was impaired due to progressive corneal oedema and glaucoma. Various considerations were taken while planning for surgical intervention because of the extensive structural damage to the anterior segment of the eye.
CONCLUSION: This case report highlights that cosmetic iris implants are dangerous intraocular devices and management of the associated complications is also challenging. As these devices cause irreversible structural and functional damage, their use should be discouraged in normal eyes.
DESIGN: Interventional case report.
METHODS: A thirty-year-old gentleman presented with complaint of progressive loss of vision after having cosmetic iris implant surgery to change his eye colour. He then developed raised intraocular pressures and had a right eye trabeculectomy conducted with the implants in situ. Subsequently, he had implant removal surgery because of persistent implant-associated complications. The vision was impaired due to progressive corneal oedema and glaucoma. Various considerations were taken while planning for surgical intervention because of the extensive structural damage to the anterior segment of the eye.
CONCLUSION: This case report highlights that cosmetic iris implants are dangerous intraocular devices and management of the associated complications is also challenging. As these devices cause irreversible structural and functional damage, their use should be discouraged in normal eyes.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app