Early fracture of the modular neck of a MODULUS femoral stem.

We present the case of a 46-year-old woman who underwent revision surgery approximately 4 years after total hip arthroplasty because of a fracture of the modular neck of a MODULUS femoral stem. The fractured surfaces of the retrieved implant were inspected using optical and scanning electron microscopy. Three-dimensional finite element analysis was also performed to identify the stresses that might have caused the failure. We concluded that active, obese patients who are implanted with a high-offset, small-sized modular component could experience stress-induced fractures of the modular neck, with proper fixation and osseointegration of the distal stem, especially if residual bone or tissue is present on the inner surface of the neck that could contribute to micromovement and decreased proximal fixation.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app