Influence of injection rates of calibrating standard solution on monitoring pulse indicator continuous cardiac output.

OBJECTIVE: This study aimed to investigate the influence of injection rates of calibrating standard solution on monitoring pulse indicator continuous cardiac output (PICCO, made in Germany), and thereby to provide significant references for clinical practice.

METHODS: A total of 108 critical patients in stroke intensive care unit were identified. All these participants received transesophageal cardiac color Doppler ultrasound, and within 15 min PICCO equipment was utilized to monitor the relevant parameters, by means of 0 °C calibrating standard solution, and the injection speeds were 2-4, 5-7, and 8-10 s. Besides, the monitoring indicators were as follows, cardiac index, global ejection fraction, global end diastolic volume index. The potential correlations were evaluated between PICCO and transesophageal cardiac color Doppler ultrasound.

RESULTS: All the data was available, and the monitored parameters of PICOO at 2-4, 5-7, and 8-10 s were positively correlated with the parameters obtained from transesophageal cardiac color Doppler ultrasound (P < 0.05). Specially, it is worth emphasizing that the best correlation between them could be provided when the injection rate was 2-4 s.

CONCLUSION: When the injection rate at 2-4 s, the parameters obtained by PICOO were much closer to that of transesophageal cardiac color Doppler ultrasound. Furthermore, the parameters of PICOO obtained at 2-4 s could better reflect cardiac function of patients.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app