We have located links that may give you full text access.
Influence of Assembly Force and Distraction on the Femoral Head-Taper Junction.
Journal of Arthroplasty 2018 July
BACKGROUND: This study investigates if the placement of femoral heads (trials and actual implants) using varying impaction forces causes physical compromise to the trunnion.
METHODS: Trunnion and head taper wear patterns were evaluated after impaction and removal of new femoral stem trunnions and ceramic heads at various impaction loads (2 kN, 4 kN, or 6 kN, n = 6/group). In addition, trunnion wear patterns were measured after plastic trials were hand-placed on new trunnions and underwent range of motion testing in a Hip Simulator (n = 5).
RESULTS: There was no significant difference in trunnion or head surface deviation, taper angle, or surface roughness in any groups preimpaction and postimpaction and removal. There was no significant surface trunnion damage from assembly and range of motion testing of the plastic femoral head trial.
CONCLUSIONS: The use of femoral head trials and the concurrent impaction and removal of a new femoral head were not associated with significant trunnion surface damage for the impaction loads observed in this study.
METHODS: Trunnion and head taper wear patterns were evaluated after impaction and removal of new femoral stem trunnions and ceramic heads at various impaction loads (2 kN, 4 kN, or 6 kN, n = 6/group). In addition, trunnion wear patterns were measured after plastic trials were hand-placed on new trunnions and underwent range of motion testing in a Hip Simulator (n = 5).
RESULTS: There was no significant difference in trunnion or head surface deviation, taper angle, or surface roughness in any groups preimpaction and postimpaction and removal. There was no significant surface trunnion damage from assembly and range of motion testing of the plastic femoral head trial.
CONCLUSIONS: The use of femoral head trials and the concurrent impaction and removal of a new femoral head were not associated with significant trunnion surface damage for the impaction loads observed in this study.
Full text links
Trending Papers
A Personalized Approach to the Management of Congestion in Acute Heart Failure.Heart International 2023
Potential Mechanisms of the Protective Effects of the Cardiometabolic Drugs Type-2 Sodium-Glucose Transporter Inhibitors and Glucagon-like Peptide-1 Receptor Agonists in Heart Failure.International Journal of Molecular Sciences 2024 Februrary 21
The Effect of Albumin Administration in Critically Ill Patients: A Retrospective Single-Center Analysis.Critical Care Medicine 2024 Februrary 8
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app