Comparison of Enzymatic Assay for HBA1C Measurement (Abbott Architect) With Capillary Electrophoresis (Sebia Minicap Flex Piercing Analyser).
Laboratory Medicine 2018 July 6
Objective: To compare the analytical performances of the enzymatic method (EM) and capillary electrophoresis (CE) for hemoglobin A1c (HbA1c) measurement.
Methods: Imprecision, carryover, stability, linearity, method comparison, and interferences were evaluated for HbA1c via EM (Abbott Laboratories, Inc) and CE (Sebia).
Results: Both methods have shown overall within-laboratory imprecision of less than 3% for International Federation of Clinical Chemistry and Laboratory Medicine (IFCC) units (<2% National Glycohemoglobin Standardization Program [NGSP] units). Carryover effects were within acceptable criteria. The linearity of both methods has proven to be excellent (R2 = 0.999). Significant proportional and constant difference were found for EM, compared with CE, but were not clinically relevant (<5 mmol/mol; NGSP <0.5%). At the clinically relevant HbA1c concentration, stability observed with both methods was acceptable (bias, <3%). Triglyceride levels of 8.11 mmol per L or greater showed to interfere with EM and fetal hemoglobin (HbF) of 10.6% or greater with CE.
Conclusion: The enzymatic method proved to be comparable to the CE method in analytical performances; however, certain interferences can influence the measurements of each method.
Methods: Imprecision, carryover, stability, linearity, method comparison, and interferences were evaluated for HbA1c via EM (Abbott Laboratories, Inc) and CE (Sebia).
Results: Both methods have shown overall within-laboratory imprecision of less than 3% for International Federation of Clinical Chemistry and Laboratory Medicine (IFCC) units (<2% National Glycohemoglobin Standardization Program [NGSP] units). Carryover effects were within acceptable criteria. The linearity of both methods has proven to be excellent (R2 = 0.999). Significant proportional and constant difference were found for EM, compared with CE, but were not clinically relevant (<5 mmol/mol; NGSP <0.5%). At the clinically relevant HbA1c concentration, stability observed with both methods was acceptable (bias, <3%). Triglyceride levels of 8.11 mmol per L or greater showed to interfere with EM and fetal hemoglobin (HbF) of 10.6% or greater with CE.
Conclusion: The enzymatic method proved to be comparable to the CE method in analytical performances; however, certain interferences can influence the measurements of each method.
Full text links
Related Resources
Trending Papers
Challenges in Septic Shock: From New Hemodynamics to Blood Purification Therapies.Journal of Personalized Medicine 2024 Februrary 4
Molecular Targets of Novel Therapeutics for Diabetic Kidney Disease: A New Era of Nephroprotection.International Journal of Molecular Sciences 2024 April 4
The 'Ten Commandments' for the 2023 European Society of Cardiology guidelines for the management of endocarditis.European Heart Journal 2024 April 18
A Guide to the Use of Vasopressors and Inotropes for Patients in Shock.Journal of Intensive Care Medicine 2024 April 14
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app