A critical evaluation of the Beckman Coulter Access hsTnI: Analytical performance, reference interval and concordance.
Clinical Biochemistry 2018 May
INTRODUCTION: We investigated the analytical performance, outlier rate, carryover and reference interval of the Beckman Coulter Access hsTnI in detail and compared it with historical and other commercial assays.
MATERIALS AND METHODS: We compared the imprecision, detection capability, analytical sensitivity, outlier rate and carryover against two previous Access AccuTnI assay versions. We established the reference interval with stored samples from a previous study and compared the concordances and variances with the Access AccuTnI+3 as well as with two commercial assays.
RESULTS: The Access hsTnI had excellent analytical sensitivity with the calibration slope 5.6 times steeper than the Access AccuTnI+3. The detection capability was markedly improved with the SD of the blank 0.18-0.20 ng/L, LoB 0.29-0.33 ng/L and LoD 0.58-0.69 ng/L. All the reference interval samples had a result above the LoB value. At a mean concentration of 2.83 ng/L the SD was 0.28 ng/L (CV 9.8%). Carryover (0.005%) and outlier (0.046%) rates were similar to the Access AccuTnI+3. The combined male and female 99th percentile reference interval was 18.2 ng/L (90% CI 13.2-21.1 ng/L). Concordance amongst the assays was poor with only 16.7%, 19.6% and 15.2% of samples identified by all 4 assays as above the 99th, 97.5th and 95th percentiles. Analytical imprecision was a minor contributor to the observed variances between assays.
CONCLUSION: The Beckman Coulter Access hsTnI assay has excellent analytical sensitivity and precision characteristics close to zero. This allows cTnI measurement in all healthy individuals and the capability to identify numerically small differences between serial samples as statistically significant. Concordance in healthy individuals remains poor amongst assays.
MATERIALS AND METHODS: We compared the imprecision, detection capability, analytical sensitivity, outlier rate and carryover against two previous Access AccuTnI assay versions. We established the reference interval with stored samples from a previous study and compared the concordances and variances with the Access AccuTnI+3 as well as with two commercial assays.
RESULTS: The Access hsTnI had excellent analytical sensitivity with the calibration slope 5.6 times steeper than the Access AccuTnI+3. The detection capability was markedly improved with the SD of the blank 0.18-0.20 ng/L, LoB 0.29-0.33 ng/L and LoD 0.58-0.69 ng/L. All the reference interval samples had a result above the LoB value. At a mean concentration of 2.83 ng/L the SD was 0.28 ng/L (CV 9.8%). Carryover (0.005%) and outlier (0.046%) rates were similar to the Access AccuTnI+3. The combined male and female 99th percentile reference interval was 18.2 ng/L (90% CI 13.2-21.1 ng/L). Concordance amongst the assays was poor with only 16.7%, 19.6% and 15.2% of samples identified by all 4 assays as above the 99th, 97.5th and 95th percentiles. Analytical imprecision was a minor contributor to the observed variances between assays.
CONCLUSION: The Beckman Coulter Access hsTnI assay has excellent analytical sensitivity and precision characteristics close to zero. This allows cTnI measurement in all healthy individuals and the capability to identify numerically small differences between serial samples as statistically significant. Concordance in healthy individuals remains poor amongst assays.
Full text links
Related Resources
Trending Papers
Heart failure with preserved ejection fraction: diagnosis, risk assessment, and treatment.Clinical Research in Cardiology : Official Journal of the German Cardiac Society 2024 April 12
Proximal versus distal diuretics in congestive heart failure.Nephrology, Dialysis, Transplantation 2024 Februrary 30
World Health Organization and International Consensus Classification of eosinophilic disorders: 2024 update on diagnosis, risk stratification, and management.American Journal of Hematology 2024 March 30
Efficacy and safety of pharmacotherapy in chronic insomnia: A review of clinical guidelines and case reports.Mental Health Clinician 2023 October
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app