3D patient imaging and retrieval analysis help understand the clinical importance of rotation in knee replacements.

PURPOSE: The purpose of the present study was to correlate highly accurate CT measurements of pre-revision total knee arthroplasty (TKA) implant position with findings of retrieval analysis post-revision, to understand the clinical relevance of TKA orientation.

METHODS: This study involved 53 retrieved TKA implants with pre-revision 3D-CT scans used to determine coronal (varus-valgus), sagittal (tibial slope) and rotational (internal rotation-external rotation) TKA orientation as well as tibiofemoral leg axis. Differences between femoral and tibial angles to describe the "relative rotational mismatch" were also calculated. All tibial inserts were forensically analyzed using the Hood score. Statistical analysis was performed to investigate correlations between TKA component orientation and surface damage (p < 0.05).

RESULTS: Femoral components were found to have axial rotations mainly within ± 3° (68%), whilst 45% of the tibial components and 66% of the relative rotational mismatches were > 3° and < - 3°, respectively. The majority of femoral and tibial components (87% in both cases), as well as the femorotibial angle (70%), showed coronal orientations within ± 3°. The 64% of the tibial components showed posterior tibial slopes out of both the 0°-3° and 5°-7° ranges. There was a significant correlation between tibial slope and damage score on polyethylene tibial inserts (r = 0.2856; p = 0.0382) as well as a significant correlation between implants' position in the axial plane and damage score on polyethylene tibial inserts (r = 0.6537, p = 0.0240).

CONCLUSIONS: This is the first study to use accurate measurements from pre-revision 3DCT to compare tibial and femoral orientation in all three planes with retrieval findings in total knee replacements. A significant correlation between implant position and polyethylene surface damage was found. These results showed the importance of optimizing component position to minimize polyethylene damage. Further analysis involving more accurate polyethylene wear measurements are fundamental to fully understand the role of components' orientation in TKAs.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app