INVESTIGATION OF A PRACTICAL PATIENT DOSE INDEX FOR ASSESSMENT OF PATIENT ORGAN DOSE FROM CONE-BEAM COMPUTED TOMOGRAPHY IN RADIATION THERAPY USING A MONTE CARLO SIMULATION.
Radiation Protection Dosimetry 2018 March 2
The purpose of this study was to investigate a practical patient dose index for assessing the patient organ dose from a cone-beam computed tomography (CBCT) scan by comparing eight dose indices, i.e. CTDI100, CTDIIEC, CTDI∞, midpoint doses f(0)PMMA for a cylindrical polymethyl methacrylate (PMMA) phantom, f(0)Ap for an anthropomorphic phantom and f(0)Pat for a prostate cancer patient, as well as the conventional size specific dose estimations (SSDEconv) and modified SSDE (SSDEmod), with organ dose for the prostate (ODprost) obtained via Monte Carlo (MC) simulation. The ODprost was the reference dose used to find the practical dose index at the center of the pelvic region of a prostate cancer patient. The smallest error rate with respect to the ODprost of 19.3 mGy (reference) among eight dose indices was 5% for f(0)Pat. The practical patient dose index was the f(0)Pat, which showed the smallest error with respect to the reference dose.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app