Impact of cleaning monitoring combined with channel purge storage on elimination of Escherichia coli and environmental bacteria from duodenoscopes.
Gastrointestinal Endoscopy 2018 August
BACKGROUND AND AIMS: We aimed to determine whether monitoring of duodenoscope cleaning by rapid adenosine triphosphate (ATP) combined with channel-purge storage could eliminate high-concern microorganisms.
METHODS: In a simulated-use study, suction channels, as well as lever recesses, from 2 duodenoscopes models and the unsealed elevator guidewire (EGW) channel from 1 of these 2 duodenoscopes (the other model has a sealed EGW) were perfused with ATS2015 containing approximately 8 Log10 colony-forming units (CFU)/mL of both Enterococcus faecalis and Escherichia coli. Pump-assisted cleaning was monitored by rapid ATP testing. Duodenoscopes exceeding 200 relative light units (RLUs) were recleaned. Clean duodenoscopes were processed through an automated endoscope reprocessor and then stored in a channel-purge storage cabinet for 1 to 3 days. Cultures of EGW channel and instrument channel combined with the lever recess (IC-LR) were taken after storage. The impacts of extended cleaning and alcohol flush were evaluated.
RESULTS: E coli was reliably eliminated in IC-LR and EGW channels of 119 duodenoscope tests (59 with sealed EGW and 60 with nonsealed EGW). However, actionable levels of E faecalis and environmental bacteria persisted. Neither alcohol flush nor extended cleaning resulted in a reduction of actionable levels for these organisms. Identification of isolates indicated that residual organisms in duodenoscope channels were hardy Gram-positive bacteria (often spore formers) that likely originated from environmental sources.
CONCLUSIONS: These data indicate that high-concern Gram-negative bacteria but not E faecalis or environmental bacteria can be reliably eliminated by use of the manufacturer's instructions for reprocessing with ATP monitoring of cleaning and channel-purge storage conditions.
METHODS: In a simulated-use study, suction channels, as well as lever recesses, from 2 duodenoscopes models and the unsealed elevator guidewire (EGW) channel from 1 of these 2 duodenoscopes (the other model has a sealed EGW) were perfused with ATS2015 containing approximately 8 Log10 colony-forming units (CFU)/mL of both Enterococcus faecalis and Escherichia coli. Pump-assisted cleaning was monitored by rapid ATP testing. Duodenoscopes exceeding 200 relative light units (RLUs) were recleaned. Clean duodenoscopes were processed through an automated endoscope reprocessor and then stored in a channel-purge storage cabinet for 1 to 3 days. Cultures of EGW channel and instrument channel combined with the lever recess (IC-LR) were taken after storage. The impacts of extended cleaning and alcohol flush were evaluated.
RESULTS: E coli was reliably eliminated in IC-LR and EGW channels of 119 duodenoscope tests (59 with sealed EGW and 60 with nonsealed EGW). However, actionable levels of E faecalis and environmental bacteria persisted. Neither alcohol flush nor extended cleaning resulted in a reduction of actionable levels for these organisms. Identification of isolates indicated that residual organisms in duodenoscope channels were hardy Gram-positive bacteria (often spore formers) that likely originated from environmental sources.
CONCLUSIONS: These data indicate that high-concern Gram-negative bacteria but not E faecalis or environmental bacteria can be reliably eliminated by use of the manufacturer's instructions for reprocessing with ATP monitoring of cleaning and channel-purge storage conditions.
Full text links
Trending Papers
A Personalized Approach to the Management of Congestion in Acute Heart Failure.Heart International 2023
Potential Mechanisms of the Protective Effects of the Cardiometabolic Drugs Type-2 Sodium-Glucose Transporter Inhibitors and Glucagon-like Peptide-1 Receptor Agonists in Heart Failure.International Journal of Molecular Sciences 2024 Februrary 21
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app