JOURNAL ARTICLE
RESEARCH SUPPORT, N.I.H., EXTRAMURAL
Add like
Add dislike
Add to saved papers

Entropic trap purification of long DNA.

Lab on a Chip 2018 March 14
Long-read genomic applications, such as genome mapping in nanochannels, require long DNA that is free of small-DNA impurities. We have developed a chip-based system based on entropic trapping that can simultaneously concentrate and purify a long DNA sample under the alternating application of an applied pressure (for sample injection) and an electric field (for filtration and concentration). In contrast, short DNA tends to pass through the filter owing to its comparatively weak entropic penalty for entering the nanoslit. The single-stage prototype developed here, which operates in a continuous pulsatile manner, achieves selectivities of up to 3.5 for λ-phage DNA (48.5 kilobase pairs) compared to a 2 kilobase pair standard based on experimental data for the fraction filtered using pure samples of each species. The device is fabricated in fused silica using standard clean-room methods, making it compatible for integration with long-read genomics technologies.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app