Intrasession Repeatability and Intersession Reproducibility Measurements Using VX120 Multidiagnostic Unit.

Eye & Contact Lens 2018 November
OBJECTIVE: The VX120 multidiagnostic unit is a multidiagnostic instrument that combines several functions: autorefraction, keratometry, corneal topography, aberrometry, pachymetry, and noncontact tonometry. The purpose of this study was to determine the intrasession repeatability and the intersession reproducibility of all parameters measured by the VX120 multidiagnostic unit in a sample of normal healthy eyes.

METHODS: Three repeated measurements in the right eye of the volunteers were taken with VX120. Repeatability of the sphere, cylinder, axis, anterior corneal powers (K flat and K steep, 3-mm pupil), high- and low-order aberration root mean square (HOA-RMS and LOA-RMS, 3-mm pupil), eccentricity, white-to-white (WTW), anterior chamber depth (ACD), and central corneal thickness (CCT) (2-mm central) was analyzed. Within-subject SD (Sw), precision, repeatability, coefficient of variation (CV), and the intraclass correlation coefficient (ICC) were calculated.

RESULTS: The CV was low for K flat and K steep, WTW, ACD, and CCT with a range from 0.34% to 1.16%. The CV was higher for sphere, cylinder, HOA-RMS, and LOA-RMS and eccentricity with a range among 6.92% to 54.24%. The ICC showed high values in all parameters except in HOA-RMS (0.720-0.776) and eccentricity (0.889) in first session with moderate agreement. Comparing the intrasession repeatability of first and second session, statistically significant differences (P<0.01) were found between both sessions just to the CV for all parameters (except cylinder values) measured with VX120. However, nonstatistically significant differences (P>0.13) were found for Sw, precision, and repeatability values.

CONCLUSIONS: The VX120 multidiagnostic unit provides repeatable measurements in anterior corneal power (K flat and K steep), WTW, ACD, and CCT. However sphere, cylinder, HOA-RMS, LOA-RMS, and eccentricity showed worse repeatability. Intersession reproducibility showed good results with little differences between sessions in healthy subjects.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app