An Inappropriate Shock Case Early after Implantation of a Subcutaneous Cardiac Defibrillator by Subcutaneous Entrapped Air.

A 17-year-old woman was resuscitated from cardiac arrest due to ventricular fibrillation and was diagnosed with concealed long QT syndrome. She underwent subcutaneous implantable cardiac defibrillator (S-ICD) implantation at our hospital. The device electrogram immediately after implantation was normal. Four days after implantation, she received an inappropriate shock. The device interrogation revealed a continuous baseline shift and frequent oversensing for low amplitude signals, followed by a shock. A chest radiograph in the orthogonal view showed entrapped subcutaneous air surrounding the distal electrode. Entrapped subcutaneous air can cause inappropriate shocks in the early period after S-ICD implantation.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app