Multicenter, multivendor phantom study to validate proton density fat fraction and T2* values calculated using vendor-provided 6-point DIXON methods.
Clinical Imaging 2018 September
PURPOSE: To evaluate the reproducibility of proton density fat fraction (PDFF) and T2* in a fat-water phantom on three different 3 T MRI systems using 6-point DIXON methods.
METHODS: A phantom which included varying fat volume percentages (true fat fraction [FF]) was scanned by three 3 T MR machines, and PDFF and T2* were measured.
RESULTS: The mean difference between true FF and PDFF was small in all vendors (-2.11% to 0.41%). However, the difference ratio for T2* values was large among vendors (1.79 to 3.36).
CONCLUSIONS: True FF and PDFF were consistent across vendors; however, T2* varied greatly.
METHODS: A phantom which included varying fat volume percentages (true fat fraction [FF]) was scanned by three 3 T MR machines, and PDFF and T2* were measured.
RESULTS: The mean difference between true FF and PDFF was small in all vendors (-2.11% to 0.41%). However, the difference ratio for T2* values was large among vendors (1.79 to 3.36).
CONCLUSIONS: True FF and PDFF were consistent across vendors; however, T2* varied greatly.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app